3 Initial Start-Up
3.1 Connection to the Mains (115 V / 230 V, 50 Hz / 60 Hz)
õ Connect the mains plug at the test instrument to the mains power out-
let. The function selector switch can be set to any position.
If a mains outlet (earthing contact outlet) is not available, or if only a
3-phase outlet is available, the adapter socket can be used to connect
the phase conductor, the neutral conductor and the protective con-
ductor. The adapter socket has three permanently attached cables
and is included with the KS13 cable set.
Attention!
!
If connection is not possible via an earthing contact outlet: Shut
down mains power first.
Then connect the cables from the coupling socket to the mains
using pick-off clips in accordance with the diagram.
Disconnection from mains power is only possible with the mains
plug.
L1
N
green-yellow
green-yellow
PE
L1
L2
L3
N
PE
L1
L2
L3
N
green-yellow
U
L–N
= 115 V / 230 V
Connection to the Mains