User Manual
GMC-I Messtechnik GmbH 9
4 Initial Start-Up
4.1 Connecting the Test Instrument to the Mains
➭ See section 12 for nominal mains values (nominal ranges of
use).
➭ Connect the test instrument to the mains cable via its inlet
plug and insert the mains plug into an electrical outlet. The
function selector switch can be set to any position.
If a mains outlet (earthing contact outlet) is not available, or if
only a 3-phase outlet is available, the adapter socket can be
used to connect the phase conductor, the neutral conductor
and the protective conductor. The adapter socket has three
permanently attached cables and is included with the KS13
cable set.
Attention!
!
If connection is not possible via an earthing contact out-
let: Shut down mains power first.
Then connect the cables from the coupling socket to the
mains using pick-off clips in accordance with the dia-
gram.
Disconnection from mains power is only possible with
the mains plug.
Figure 1 Connecting the Test Instrument to the Mains
4.2 Automatic Recognition of Mains Connection Errors
The device automatically recognizes mains connection errors if
the conditions in the following table have been fulfilled. The user is
informed of the type of error, and all measuring functions are dis-
abled in the event of danger.
1
10 A R
PE
measurements are only possible with line voltages of 115/230 V and line
frequencies of 50/60 Hz.
Attention!
!
If, while testing protective conductor potential, you deter-
mine that the mains protective conductor is conducting volt-
age (in accordance with the first two cases listed above),
no further measurements may be performed with the test in-
strument. If this is the case, potentially dangerous voltage
is also present at the accessible earthing contacts of the
standard socket (test socket). Immediately disconnect
the test instrument from the mains and arrange to have
the fault eliminated at the mains connection.
Note
Voltage at the electrical system’s protective conductor PE may
result in distorted measurement values during testing for
the absence of voltage, or during leakage voltage mea-
surements.
4.3 Test Probe Connection (P1)
Insert the double plug from test probe P1 into socket 1 such that
the plug with the white ring makes contact with the socket with
the vertical bar.
The white ring identifies the terminal for the high current conduc-
tor which is safeguarded by the neighboring fuse link.
Note
Difficultly in contacting exposed conductive parts when using
the standard probe with test tip
In order to assure good contact, surface coatings must
be removed from devices under test with special tools at
a suitable location.
The tip of test probe P1 is not suitable for scratching
away paint, because this may impair its coating and/or
mechanical strength. Brush probe Z745G may be more
suitable than the test probe in certain individual cases.
L1
N
green-yellow
green-yellow
PE
L1
L2
L3
N
PE
L1
L2
L3
N
green-yellow
KS13
Type of Mains
Connection Error
Message Condition Measurements
Voltage at
protective conductor PE
to finger contact
(START/STOP key)
Display at the
instrument
Press START/STOP
button
U 25 V
All measurements
disabled
Protective conductor PE
and phase conductor L
reversed and/or
neutral conductor N
interrupted
Voltage at PE
> 100 V
Impossible
(no supply power)
Line voltage
< 180 V / < 90 V
(depending on mains)
Reference
voltage
adjustable
in setup
U
L-N
< 180 V
U
L-N
<90V
(reference
voltage adjustable
in setup)
Conditionally
possible
1