User Manual
42 GMC-I Messtechnik GmbH
Sequence Parameters (as from Firmware V1.3.0)
The sequence parameters of switch positions A1 ... A8 and
AUTO allow for the following settings and are the same for all
switch positions:
• Showing and hiding of selected test steps
• Selection of test current for RPE
• Measurements with or without polarity reversal
• Showing or hiding of test comments
10.3 Connecting the DUT
➭ Connect the DUT to the test instrument in accordance with
the selected test sequence.
– Test socket
– Permanent connection
– AT16/32-DI adapter
Switch position: A1 ... A7, AUTO
Connection depends on the type of DUT.
Switch Position A8
For testing extension cords in accordance with standards: con-
nection to the test socket via the following adapter:
– EL1:
for single-phase extension cords
10.4 Selecting a DUT
➭ If no DUT has been selected in the initial display, enter its ID
number, for example by means of a barcode scanner, after
selecting ID.
➭
As an alternative
, activate the database view with the
MEM key:
➭ Select the DUT for the test sequence with the scroll keys.
➭ Return to the measuring view by pressing the MEM key.
10.5 Checking Connection and Starting the Test Sequence
➭ Trigger the connection test and the test sequence
by pressing the START key.
The following checks are run automatically before the
test sequence is started:
• Probe test (whether or not test probe P1 is connected)
• Insulation test (whether or not the DUT is set up in a well-insu-
lated fashion)
• On test and short-circuit test:
(Prerequisite: sequence parameter „short-circuit check L-N“
has been preset to „on“)
In order to be able to detect a short-circuit at the DUT, testing
is conducted between L and N, as well as LN and PE.
If you have set the “
Detected classification
” parameter for the
respective test sequence to “Always accept” and the “
Auto-
detection of
” parameter to “
Connection and PC
” (before trig-
gering Start), the following additional checks will be run
before the test sequence is started:
•
Protection class detection for DUTs with protective conductor
• Connection check: whether or not the DUT is connected to
the test socket. In the case of protection class I: whether or
not the two protective conductor terminals are short-circuited.
Sequence parameter Setting options / Meaning
RPE Test step protective conductor resistance
on: perform test step
off: hide/skip
RINS PCI+II Test step insulation resistance Protection class I and II
on: perform test step
off: hide/skip
RINS PCIII (SELV) Test step insulation resistance Protection class III
and/or safety extra low voltage (SELV)
on: perform test step
off: hide/skip
RPE IP Select test current
200 mA~
10 A~ (only SECUTEST B
ASE10 or SECUTEST XTRA)
IPE measurement type
(active)
direct or differential
Visual inspection on: perform at the start of the test sequence
off: hide/skip
Function test on: perform at the end of the test sequence
off: hide/skip
Reverse polarity on: perform measurements in both directions
off: perform measurements without polarity reversal
Show test comments on: show test comments
off: hide test comments
Short-circuit check L-N on: perform short-circuit detection
off: hide/skip short-circuit detection
MEM