GMC-I Messtechnik GmbH 41
Meaning of Symbols in the User Interface – Test Sequence 10.2 Selecting and Configuring a Test Sequence
Initial Page of a Sample Test Sequence
Classification Parameters – Automatic Recognition
If the settings of certain classification parameters are auto-
matically recognized by the test instrument, it is signalled by an
amber frame (as from firmware version V1.3.0; here: connection
type test socket and protection class I). The description of these
parameters, which depend on the switch position, is given below.
Classification Parameters, Switch Settings A1 through A8
Classification Parameters, AUTO Switch Setting
* These parameters must be entered manually if they’re not automatically
detected, or if they’re detected incorrectly.
Sym-
bol
Softkey Variants, Test Sequence
Test for Protection Class I Devices
Exposed, conductive parts are connected to the protec-
tive conductor so that they are not charged with voltage if
the basic insulation should fail.
Test for Protection Class II Devices
These devices are equipped with double insulation or
reinforced insulation.
Test for Protection Class III Devices
These devices are supplied with safety extra-low voltage
(SELV). Beyond this, no voltages are generated which
exceed SELV.
Adjust measuring parameters.
Currently no function – for future expansion
Set classification parameters
Assess visual inspection or function test with OK
or not OK (toggle key)
Enter a comment, e.g. for the visual inspection or function
test
Edit a comment
Continue test, next test step in the test sequence
Accept changed parameter, return to memory view
Stop test sequence
Skip measurement
Start evaluation – record measured value. Each time this
softkey is pressed, an additional measured value is saved
and the number is increased by one.
Repeat measured value recording
Delete measured value
Display measured values
Display details from the results list
Hide details from the results list
The ID number to which the measurement(s) will be stored
can be entered here.
Valid measured values have been obtained for a test
sequence. This measurement can be saved.
Save measurement data as (with display of directory path
/ ID or new entry of an ID other than the preselected one)
Sequence Parameter Setting Options / Meaning
1/2
Standard Permanently set
Protection class * Selection of PC1, PC2, PC3 or combinations
Connection type * Test socket / permanent / adapter
2/2
Measurement type (MA) Permanently set
Detected classification No auto-detection:
All classification parameters such as connection and protec-
tion class must be entered manually.
Always accept:
All classification parameters activated under “Auto-detection
of” are detected automatically and accepted.
Auto-detection of Any desired combinations for automatic detection of:
– Connection
– Protection class (PC)
Sequence Parameter Setting Options / Meaning
1/2
Standard Test standard / extension cord
Protection class * PC1/PC2/PC3
Connection type * Test socket / permanent / adapter
2/2
Measurement type (MA) Active or passive DUT
(on test:
on = passive, off = active)
Detected classification No auto-detection:
All classification parameters such as connection and protec-
tion class must be entered manually.
Always accept:
All classification parameters activated under “Auto-detection
of” are detected automatically and accepted.
Auto-detection of Any desired combinations for automatic detection of:
– Connection
– Protection class (PC)
A1
Standard Connection Type Protection Class
Class. Parameters
Meas. Type
Status Line
Sequ. Parameters