User Manual

GMC-I Messtechnik GmbH 27
Test Sequence for Alternative Measuring Method
Set the rotary switch to the I
PE
position.
Select the Alternative measurement type:
– By setting the parameters
or
– Via the MA key
Set the Up(set) and frequency parameters.
Connect the DUT’s mains plug (protection class I) to the test
instrument’s test socket (see figure 4/4).
Start the test: press the START/STOP key.
Switch the device under test on.
The measured values are displayed. The measured
value recording symbol shown at the right appears.
Each time this key is pressed, the currently displayed
measured value is saved to the buffer.
Stop the test: press the START/STOP key.
The save symbol appears (floppy disk showing the
number of measured values stored to buffer memory)
and prompts you to save the measured values to an ID
number.
Read the measured values and compare them with the
table of permissible limit values.
Press the ESC key in order to discard the measured
values stored to buffer memory and acknowledge by
pressing the key shown at the right.
Maximum Permissible Limit Values for Leakage Current in mA
* For devices with heating power of greater than 3.5 kW
Note 1:Devices which are not equipped with accessible parts that are connected to
the protective conductor, and which comply with requirements for touch cur-
rent and, if applicable, patient leakage current, e.g. computer equipment with
shielded power pack
Note 2:Permanently connected devices with protective conductor
Note 3:Portable x-ray devices with mineral insulation
Key
I
PE
Current in the protective conductor (primary leakage current)
Test Standard I
PE
VDE 0701-
0702:2008
PC I: 3.5
1 mA/kW *
DIN EN 60974-4
VDE 0544-
4:2009-06
5 mA