Meaning of Symbols in the User Interface –
Single Measurements
8.2 Measurement Series and Storage
Single measurements can be combined into measurement series.
The measured values can be saved by pressing the save key, or
measurement series can be generated. These can be saved to a
DUT (ID number) which has already been set up in the database
(see section 5.4.1). The appearance of the save key changes
depending on meaning.
Measuring Sequence with Pre-Selection of the DUT
1 Activate the database view (MEM navigation) by pressing the
MEM key.
2 Select the DUT or its ID number for the following measure-
ments with the scroll keys.
3 Return to the measuring view by pressing the MEM key or
the START/STOP key.
4 Start the test with the START/STOP key.
The symbol shown at the right appears and the zero
indicates that no measurements have yet been
recorded or saved to buffer memory.
5 Each time the key at the right is pressed, the respec-
tively current measured value is saved to buffer mem-
ory and the number shown in the symbol is increased.
In this way, you always know how many measurements have
already been recorded.
6 Stop the measurement by pressing the START/STOP
key, unless a specified measuring time has been stipu-
lated.
The Save as symbol appears (floppy disk icon with the
number of measured values saved to the clipboard).
7 If you press the save symbol now (floppy disk), the dis-
play is switched to the DUT in the database view for checking.
8 After pressing the save symbol once again, acknowledgement
of successful storage appears.
At the same time, the display is switched to the measuring
view.
Measuring Sequence with Subsequent Entry of the DUT
1 Start the measurement by pressing the START/STOP but-
ton.
The symbol shown at the right appears and indicates
how many measurements have already been performed.
2 Stop the measurement by pressing the START/STOP
key, unless a specified measuring time has been stip-
ulated.
The save symbol (floppy disk with a number 1)
appears and indicates that one valid measured value has
been recorded, which can now be saved.
3 Press the save symbol (floppy disk).
4 You are informed that you haven’t selected a DUT in
the database.
5 There are two ways to subsequently select a DUT
using an ID number which has already been set up in
the database:
– Select the ID number with a barcode scanner
or
– Enter an ID number by pressing the ID key.
6 The cursor jumps to the location of the DUT with the selected
ID number. You only need to acknowledge this position by
pressing the green checkmark.
7 Press the save symbol (floppy disk).
A message appears indicating that the data have been suc-
cessfully saved and the display is switched to the measuring
view.
Note
If the entered number cannot be found in the database
(because it hasn’t been set up), it can be entered imme-
diately by pressing Yes when the prompt appears.
However, the storage location cannot be selected in this
case. The measurement is saved to the most recently
selected hierarchy.
Note
Measurements and measurement series can only be
saved after measurement has been completed. Mea-
sured values can only be added to intermediate buffer
memory during a measurement. Customer, location and
other entries cannot be changed in the memory menu.
These have to be selected directly in the database and
entered or changed.
Sym-
bol
Softkey Variants, Single Measurements
Set parameters
Accept changed parameters, acknowledge memory loca-
tion
Acknowledge messages during tests/measurements
or resume test sequence
Abort measurement
Direct selection key for selecting the measurement type
Currently selected polarity: “L-N”
Press key to change polarity
Currently selected polarity: “N-L
Press key to change polarity
Ip
Direct selection key for selecting test current for protective
conductor measurement
U+
U–
Direct selection key for changing voltage in 10 V steps for
insulation measurement
Start evaluation – record measured value. Each time this
softkey is pressed, an additional measured value is saved
and the number is increased by one.
The ID number to which the measurement(s) will be stored
can be entered here.
Valid measured values have been obtained for a measure-
ment. This measurement can be saved.
Save measurement data as (with display of directory path
/ ID or new entry of an ID other than the preselected one)
0
1
3
3
3