User Manual

10 GMC-I Messtechnik GmbH
4.4 Device Settings
For the purpose of initial start-up, we recommend
setting the following basic parameters in the order
shown at the right:
Setup 2/3 > Culture > Language (for user interface)
Setup 2/3 > Culture > Keyboard Layout (for alphanumeric entries)
Setup 1/3 > System > Date / Time (for reports generating)
Setup 1/3 > System > Brightness (display brightness as %)
Figure 2 Device Settings, Main Menu Level – SETUP Switch Setting
The following parameters are advisable for maintenance purposes:
SETUP 3/3 > Test > Display / Buzzer (for checking info and warning
displays/signals)
SETUP 3/3 > System info > Software version for updates and cali-
bration data for recalibration
See section 13.3 regarding downloading the latest software ver-
sion.
Display additional menu pages
Menu selection for operating parameters page 1 of 3
Parameters for single meas. and test sequences
Set test sequence parameters
Clear database or display statistics,
Set date/time, volume, and brightness,
Display additional menu pages
Menu selection for operating parameters, page 2 of 3
To submenu for selecting and setting up printer
Select inspector or enter a new one
Select language for user interface and keyboard
For example, select and config. barcode scanner
Display additional menu pages
Menu selection for operating parameters, page 3 of 3
Self-test for display and buzzer
Query software/hardware version,
see submenu on following page
Keys with fixed functions Softkeys = keys with functions that vary
depending on the operating level
PRINT
ESC
HELP
MEM
PRINT
ESC
HELP
MEM
PRINT
ESC
HELP
MEM
Setup 1/3
Setup 2/3
Setup 3/3
See following page
serial number and calibration data
Save
only appears if a USB stick is plugged in
by scanning the displayed barcode