Datasheet

SECUTESTBASE, BASE10 and XTRA
Test Instruments for
Measuring
Electrical Safety of Devices
6 GMC-I Messtechnik GmbH
Influencing Quantities and Influence Error
Reference Ranges
Line voltage 230 V AC 0.2%
Line frequency 50 Hz 2Hz
Waveform
Sine (deviation between effective and rectified value < 0.5%)
Ambient
temperature +23 C 2K
Relative humidity 40 60%
Load resistance Linear
Nominal Ranges of Use
Nominal line voltage 100 V 240 V AC
Nominal line frequency50 Hz ... 400 Hz
Line voltage
waveform Sinusoidal
Temperat ure 0 C +50C
Ambient Conditions
Storage temperature 20 C +60C
Operating temperature 5 C +40C
Accuracy range 0 C +40C
Relative humidity Max. 75%, no condensation allowed
Elevation Max. 2000 m
Deployment
Indoors, except within specified ambient
conditions
Power Supply
Electrical system
TN or TT
Line voltage 100 V 240 V AC
Line frequency 50 Hz ... 400 Hz
Power consumption 200 mA test: approx. 32 VA
10 A test: approx. 105 VA
For function test Continuous max. 3600 VA, power is con-
ducted through the instrument only,
switching capacity 16 A, ohmic load
Electrical Safety
Protection class I per IEC 61010-1/EN 61010-1/
VDE 0411-1
Nominal voltage 230 V
Test voltage 2.3 kV AC 50 Hz or 3.3 kV DC
(mains circuit / test socket to mains PE ter-
minal, USB, finger contact, probe, test
socket)
Measuring category 250 V CAT II
Pollution degree 2
Safety shutdown
At DUT differential current of > 10 mA,
shutdown time: < 100 ms,
can also be set to > 30 mA
with following probe current during:
– Leakage current meas.:
> 10 mA~/< 5 ms
– Protective conductor resistance meas.:
>250mA~/<1ms
Fuse links Mains fuses: 2 ea. 500 V/16 A FF
Probe fuse: 250 V/250 mA MT
SECUTEST B
ASE10: Additionally
1 ea. 500 V/16 A FF
Electromagnetic Compatibility
Product standard DIN EN 61326-1
USB Data Interface
Type USB slave for PC connection
Type 2 ea. USB master for external keyboard,
for barcode reader,
for USB stick for data backup,
for printer
Mechanical Design
Display 4.3" multi-display (9.7 x 5.5 cm),
backlit, 480 x 272 pixels at 24 bit color
depth (true color)
Dimensions W x H x D: 295 x 145 x 150 mm
Height with handle: 170 mm
Weight Approx. 2.5 kg
Protection Housing: IP 40
Test socket: IP 20 per DIN VDE 0470,
part 1/EN 60529,
Table Excerpt Regarding Significance of IP
Codes
Influencing Quantity /
Sphere of Influence
Designation
per
DIN VDE 0404
Influence Error
% rdg.
Change of position E1
Change to test equipment supply
voltage
E2 2.5
Temperature fluctuation
E3
Specified influence error valid
starting with temperature changes
as of 10 K:
0 40
C2.5
Amount of current at DUT E4 2.5
Low frequency magnetic fields E5 2.5
DUT impedance E6 2.5
Capacitance during insulation mea-
surement
E7 2.5
Waveform of measured current
E8
49 51 Hz
2 with capacitive load (for equiva-
lent leakage current)
45 100 Hz 1 (for touch current)
2.5 for all other measuring ranges
Interference
Emission
Class
EN 55011 B
Interference
immunity
Test value Evaluation criterion
EN 61000-4-2 Contact/atmos. – 4 kV/8 kV A
EN 61000-4-3 3 V/m or 1 V/m A
EN 61000-4-4 1 kV B
EN 61000-4-5 1 kV or 2 kV A
EN 61000-4-6 3 V/m A
EN 61000-4-11 0.5/1/25 periods A
250 periods C
IP XY
(1
st
digit X)
Protection Against
Foreign Object Ingress
IP XY
(2
nd
digit Y)
Protection Against
Penetration by Water
2 12.5 mm dia. 0 Not protected
4 1.0 mm dia. 0 Not protected