Datasheet

GMC-I Messtechnik GmbH 5
SECUTESTBASE, BASE10 and XTRA
Test Instruments for
Measuring
Electrical Safety of Devices
Characteristic Values
1
Specified values are only valid for the display at the test instrument. Data transmit-
ted via the USB port may deviate from these values.
2
Known as equivalent leakage current or equivalent patient leakage current from previ-
ous standards
3
Protective conductor current, touch current, device leakage current, patient leak-
age current
4
Protective conductor current, touch current, device leakage current
5
Only with SECUTEST BASE10 and SECUTEST XTRA with feature G01
Key: rdg. = reading (measured value), d = digit(s)
Test Times, Automated Sequence
Test time > 2 s, exception: device protective conductor resistance
R
PE: > 7 s. Test times are not checked or calibrated, but rather
determined on the basis of processor cycle times.
Emergency Shutdown During Leakage Current Measurement
As of 10 mA of differential current (can also be set to 30 mA),
automatic shutdown ensues within 100 ms.
Func-
tion
Measured
Quantity
Meas. Range /
Nominal Range of
Use
Reso-
lution
Nominal
Voltage
U
N
Open-
Circuit
Voltage
U
0
Nom.
Current
I
N
Short-
Circuit
Current
I
K
Inter-
nal Re-
sis-
tance
R
I
Refer-
ence
Resis-
tance
R
REF
Measuring
Uncertainty
1
Intrinsic Error
1
Overload
Capacity
Value Time
Tests, 62638 (DIN VDE 0701-0702) / IEC 62353 (VDE 0751)
Protective
conductor
resistance
R
PE
000 999 m 1m
< 24 V
AC or DC
>200 mA
AC or DC
> 10 A AC
5)
——
(15% rdg.+ 10 d)
>10d
(10% rdg.+ 10 d)
>10d
264 V
250 mA
16 A
5)
Cont.
1.00 ... 999 10 m
10.0 30.0
100 m
Insulation
resistance
R
ISo
10 999 k 1k
50 500
VDC
1.0 • U
N
1.5 • U
N
>1 mA >2mA
(5% rdg.+ 4 d)
> 10 d
(2.5% rdg.+2 d)
> 10 d
264 V Cont.
1.00 9.99 M 10 k
10.0 99.9 M 100 k
20 M:
(10% rdg.+ 8 d)
20 M:
(5% rdg.+4 d)
100 300 M 1M
Leakage current,
alternative
measurement
2
IPE, IB, IG, IA
0.0 ... 99 A1A
50 ...
250 V~
20/+10%
—>1.5mA
> 150 k
1k
10
(5% rdg.+ 4 d) > 10 d
> 15 mA:
(10% rdg.+ 8 d)
(2% rdg.+2 d) > 10 d
> 15 mA:
(5% rdg.+ 4 d)
264 V Cont.
100 ... 999 A1A
1.00 ... 9.99 mA 10 A
10.0 ... 30.0 mA 100 A
Leakage current,
direct
measurement
3
IPE, IB, IG, IA, IP
Only Ip: 0.0 ...
99.9 A
100 nA
——
1k
10
(5% rdg.+ 4 d)
> 10 d
(2.5% rdg.+2 d)
> 10 d
264 V Cont.
0.0 ... 99 A1A
100 ... 999 A1A
1.00 ... 9.99 mA 10 A
10.0 ... 30.0 mA 100 A
Leakage current,
differential
current
measurement
4
IPE, IB, IG
0 ... 99 A1A
——
1k
10
(5% rdg.+ 4 d)
> 10 d
(2.5% rdg.+2 d)
> 10 d
264 V Cont.
100 ... 999 A1A
1.00 ... 9.99 mA 10 A
10.0 ... 30.0 mA 100 A
Function test
Line voltage
U
L–N
100.0 240.0 V~ 0.1V ————
(2% rdg.+2 d)
264 V Cont.
Load current I
L
0 16.00 A
RMS
10mA ————
(2% rdg.+2 d)
16 A Cont.
Active power P
0 3700 W 1 W
(5% rdg.+10 d)
>20d
264 V Cont.
20 A 10 min
Apparent power S
0 4000 VA 1 VA Calculated value, U
L–N
• I
V
(5% rdg.+10 d)
>20d
Power factor PF
with sinusoidal
waveform: cos
0.00 1.00 0.01 Calculated value, P / S, display > 10 W (10% rdg.+5 d)
UP
ROBE
Probe voltage
(phase search)
, and
0.0 99.9 V 100 mV
——
(2% rdg.+2 d)
264 V Cont.
100 ... 300 V 1 V
t
A
PRCD
Time to trip at 30 mA 0.1 ... 999 ms 0.1 ms
5 ms