Operation Manual

GMC-I Messtechnik GmbH 77
Automatic Test – Individual Test Sequences
6.3.3 “Sequences” Index Card
All of the test sequences entered with the test sequence adminis-
tration function (section 7.4) are listed here, as long as no prese-
lection has been entered for a customer, a location or an object in
the “Objects” index card (see page 76).
If you select a device under test, only those sequences are dis-
played which have been assigned to the DUT. Any desired test
sequence can be started by pressing and holding the ONSTART
key.
6.3.4 “Test” Index Card
Selection
The desired device under test can be loaded by entering the ID
number or by scanning. Preset filters are disabled in this case,
and are overwritten by the data associated with the entered
device under test.
Test Sequence
Start the test with the ONSTART key.
Checking is first executed in order to determine whether a sched-
uled test or a spontaneous test will be conducted.
If a spontaneous test is conducted, this is documented later in the
test results under activity.
Depending upon the sequence, the corresponding tests are
executed (e.g. special test, short-circuit, potential at protective
conductor etc.). All test steps of the sequence are successively
executed.
An hourglass appears in the “Test” index card.
As long as the test is being executed, the testing ... icon is