Equipment Leakage Current
Sequence
Ð Select the test: I EL softkey.
Ð Connect the DUT to the test socket, or connect the probe.
Ð Select measuring range, test voltage Ut, test frequency ft and
measurement duration.
Ð Select type of connection and type of current via measurement
type.
Ð Start the test: Press the ONSTART key.
Ð After each reconnection to the mains, and as soon as the first
test is started, a mains connection test is executed
(see section 4.1.1).
Ð Measured values are displayed.
Ð End the test: Press the ONSTART key.
Ð Read the measured value and compare it with the table of
allowable limit values.
Maximum Allowable Limit Values for Equipment Leakage Current in mA
I
EL
Equivalent leakage current
I
EHL
Equivalent housing leakage current
PE Protective conductor
1)
For devices with heating power of greater than 3.5 kW
2)
This limit value is not taken into consideration in the IEC 62353
(VDE 0751-1) standard.
Test Standard I
EL
I
EDL
VDE 0701-0702
SC I: 3.5 / 1 mA/kW
1)
SC II: 0.5
VDE 0751: 2008
SC II
0.2
2)
SC I (PE or parts connected to PE)
1
Permanently connected devices with PE
10
Portable x-ray devices with additional PE
5
Portable x-ray devices without additional PE
2
Devices with mineral insulation
5