Operation Manual
142 GMC-I Messtechnik GmbH
Appendix
Medical devices without application parts
0751SK I 0751 SK II
Visual inspection Visual inspection
Contact exposed conductive parts of the
DUT which are connected to the protec-
tive conductor (PE) with the probe
(socket 1/2).
R-PE / Pd 0,25 / GW 0,3
Switch the device under test on. Switch the device under test on.
Contact exposed conductive parts
of the DUT
with the probe (socket
1/2).
Attention! Mains voltage will be ap-
plied to the DUT.
Attention! Mains voltage will be ap-
plied to the DUT.
I-GA (L/N) / Diff 230 / GW 0,5 mA
I-B (L/N) / Direkt 230 / GW 0,1 mA
Attention! Mains voltage will be ap-
plied to the DUT.
Attention! Mains voltage will be ap-
plied to the DUT.
I-GA (N/L) / Diff 230 / GW 0,5 mA
I-B (N/L) / Direkt 230 / GW 0,1 mA