Operation Manual
80 GMC-I Messtechnik GmbH
Automatic Test – Individual Test Sequences
6.3.6 Evaluation of Measured Values for Individual Test Steps
The evaluation of the test steps is based upon limit values which have been stipulated in the respective test step specifications (see
section 7.4.2). This includes the specification of a lower and/or an upper limit value and of the setpoint, if applicable. The respectively
poorest results with reference to the specified limit value or the setpoint are displayed and saved.
Test OK/passed:
The measured value with the smallest difference relative to
the MinLimitValue is displayed and saved, in this case MV3.
Test not passed:
The measured value with the largest difference relative to
the MinLimitValue is displayed and saved, in this case MV1.
Test OK/passed: The measured value with the largest difference
relative to the setpoint is displayed and saved, in this case MV7.
Test not passed:
The measured value with the largest difference rel-
ative to the MinLimitValue is displayed and saved, in this case
MV1.
Test OK/passed:
The measured value with the smallest difference relative to
the MaxLimitValue is displayed and saved, in this case MV5.
Test not passed:
The measured value with the largest difference relative to
the MaxLimitValue is displayed and saved, in this case MV7.
Test OK/passed: The measured value with the largest difference rel-
ative to the setpoint is displayed and saved, in this case MV1.
Test not passed:
The measured value with the largest difference rel-
ative to the MaxLimitValue is displayed and saved, in this case
MV7.
Test step spec.
MinLimitValue l
Setpoint —
MaxLimitValue —
Test step spec.
MinLimitValue l
Setpoint l
MaxLimitValue —
MinLimitValue
MV2 MV3
MV4 MV6
MV1
MV7
OK
A
MV5
SetpointMinLimitValue
MV2 MV3
MV4 MV6
MV1
MV7
OK
B
MV5
Test step spec.
MinLimitValue —
Setpoint —
MaxLimitValue l
Test step spec.
MinLimitValue —
Setpoint l
MaxLimitValue l
MaxLimitValue
MV2 MV3
MV4 MV6
MV1
MV7
OK
A
MV5
Setpoint MaxLimitValue
MV2 MV3
MV4 MV6
MV1
MV7
OK
A
MV5