Operation Manual
76 GMC-I Messtechnik GmbH
Automatic Test – Individual Test Sequences
6.3.2 “Objects” Index Card
All of the DUTs entered with the object administration function
(section 7.3) are listed here, provided no pre-selection has been
made for object type, customer or location, and no scheduled
tests have been selected in the “Deadlines” index card (“filter
function”, see page 74).
You can selected a device under test from the list or by entering
its ID number. The pertinent data for the object including
designation, object type, customer and location are displayed
automatically. If a deadline is associated with the object which
should not be executed on the specified date, a selection must be
made:
– New activity: The following test is deemed executed on the
future deadline and no longer appears in the deadlines list
after execution and storage of test results.
The “New activity” designation can be changed – Shown in
the next screenshot as “Periodic Test”. See section 4.11.9
– Spontaneous test: The following test is not deemed executed
on the future deadline and thus still appears in the deadlines
list.
Filter:
Editor:
edit object data
enter preselection