Insulation Resistance
SC1 with Mains Plug
DUT mains plug connected to the test socket
SC2 with Mains Plug
DUT mains plug connected to the test socket
Probe connected to sockets 1 and 2
Extension Cable
Extension cable plug connected to the test socket
Connection of Permanently Installed Safety Class I Devices
Attention!
!
Deactivate the electrical system which supplies power to
the device under test before connecting the test system!
Ð Remove the mains fuses from the device under test and
disconnect the neutral conductor N inside the device
under test.
Ð Connect the probe to phase conductor L at the device under
test in order to measure insulation resistance.
Permanent Connection
Probe connected to sockets 1 and 2