Operation Manual
40 GMC-I Messtechnik GmbH
Manual Test Individual Measurement
5 Manually Triggered Measurements –
Individual Measurements
Individual tests are executed at the “Manual Test” menu level.
Tests can be executed here in a targeted fashion while repairing
defective devices under test.
Other applications are started in the accessories or setup menu.
You should double check all parameters before starting the re-
spective measurement with the ONSTART key. Adjustable param-
eters are highlighted by means of raised buttons. Before starting
the actual measurement, it is also advisable to conduct a visual
inspection, as well as a connection test with the help of the test
function.
5.1 General Procedure
Select the “Manual Test” main menu .
Ð Select the measuring function.
Ð Double check the selected parameter settings.
Readjust them if necessary.
Ð Select the desired type of measurement: type of DUT, type of
connection.
Ð Connect the DUT in accordance with the wiring diagrams
provided by the help function (HELP button) and the above
selected type of connection.
Depending upon the type of connection selected for the DUT, use
of probes, adapters or a clip-on ammeter may be required.
Ð Start the test with the ONSTART key.
The following symbol appears for the duration of the
measurement:
After measurement has been completed – after the selected
measuring time has elapsed, or if (long-term) measurement is
aborted by pressing the ONSTART key – the following symbol ap-
pears:
Measurement data can be read from the display or printed out.
Ð The display is returned to the main menu by pressing the ESC
or the MENU key.
Man Test
MENU
or