Operation Manual
34 GMC-I Messtechnik GmbH
Initial Start-Up – Setup
Automatic Update Sequence
1 Update preparation: advancing blue bar
2 Operating system is rebooted: initial screen with an advancing blue bar –
pause for approx. 30 seconds
3 The update is loaded (approx. 7 min.): two advancing blue bars
4 Operating system is rebooted: initial screen with an advancing blue bar
5 Calibration prompt
6 Login window appears
7 The software update has been completed and the CF card can be removed
after pressing the button at the left-hand side of the slot.
Beachten Sie Folgendes bei einem Update von Version 1.8.2 auf
Version 1.8.3 oder höher (nur SECUSTAR FM):
As of version 1.8.2, the test system has several hierarchies. For this
reason, you’ll need an LHTIF.exe file (version >= 1.05.0013) or
PS3 as of version 9.10.0113 in order to exchange data with PS3.
All of the necessary files can be downloaded from our website at
http://www.gossenmetrawatt.com/deutsch/produkte/secustar-up.htm
.
It’s important to sort data into the correct hierarchies when
updating.
If you’re working with PS3 (lower than version 9.10.0.113) or
LHTIF.exe (version >= 1.05.0013) and test system firmware version
1.8.2, and if data are included in the test system in the HK11
hierarchy, they must be transferred to hierarchy HK4.
Proceed as follows to this end:
Ð Back your data up to an XML file.
Ð Open the XML file in an editor.
Ð Replace all occurrences of hierarchy-ref=“11“ in the text with
hierarchy-ref=“4“.
Ð Save the XML file.
Ð Run DB Init at the test system.
Ð Import the modified XML file.