Operation Manual

GMC-I Messtechnik GmbH 155
Maintenance – Calibration
Data Security
Error message during creation of the auto-backup
file (filename.bak)
The CF card is not inserted.
The CF card does not have enough memory capac-
ity.
Insert the CF card into the device.
Reduce the backup versions.
Skip the auto-backup function by clicking “Ignore”.
Error message during automatic creation the backup
file (filename.xml) before an update
The CF card is not inserted.
The CF card does not have enough memory
capacity.
Insert a CF card with adequate memory capacity into
the device.
Execute a manual XML export in order to receive an
exact error message (see section 7.7.6).
Error message during creation of an update
The battery pack has not been inserted and
connected.
The battery pack is not adequately charged.
Charge the battery pack, insert it and connect it.
Compact flash card cannot be read You accessed the card to early.
After inserting the CF card, wait for 10 to 15 seconds
until the system has detected and loaded the card.
The test system slows down. Too many data records at the test system
Use an external PS3 / visual FM database to manage
your data, and load only those data required for the
test to the test system.
Displays and Error Messages Meaning / Cause Remedy