Operation Manual

140 GMC-I Messtechnik GmbH
Appendix
Medical devices with type B application parts
0751 SkI B 0751 SK II B 0751 SKI + II B
Visual inspection Visual inspection Visual inspection
Contact exposed conductive parts of the DUT which are
connected to the protective conductor (PE) with the probe
(socket 1/2).
Contact exposed conductive parts of the DUT
with the
probe (socket 1/2).
Contact exposed conductive parts of the DUT which are
connected to the protective conductor (PE) with the probe
(socket 1/2).
R-PE / Pd 0,25 / GW 0,3 R-PE / Pd 0,25 / GW 0,3
Contact the probe (socket 1/2) with all exposed conductive
parts and the application parts of the DUT.
Attention! Mains voltage will be applied to the DUT. Attention! Mains voltage will be applied to the DUT. Attention! Mains voltage will be applied to the DUT.
I-GA (L/N) / Diff 230 / GW 0,5 mA
Contact type B application parts at the DUT with the probe
(socket 1/2).
Contact exposed conductive parts of the DUT which are
not connected to the protective conductor (PE) with the
probe (socket 1/2).
I-GA (L/N) / Diff 230 / GW 0,5 mA
I-B (L/N) / Direkt 230 / GW 0,1 mA I-B (L/N) / Direkt 230 / GW 0,1 mA
Contact exposed conductive parts of the DUT
with the
probe (socket 1/2).
Contact the type B application parts at the DUT with the
probe (socket 1/2).
I-PA(L/N) Direkt /230 / GW 0,1 mA I-PA(L/N) Direkt /230 / GW 0,1 mA I-PA(L/N) Direkt /230 / GW 0,1 mA
Contact exposed conductive parts of the DUT
with the
probe (socket 1/2).
Contact the probe (socket 1/2) with all exposed conductive
parts and the application parts of the DUT.
Attention! Mains voltage will be applied to the DUT. Attention! Mains voltage will be applied to the DUT. Attention! Mains voltage will be applied to the DUT.
I-GA (N/L) / Diff 230 / GW 0,5 mA I-GA (N/L) / Diff 230 / GW 0,5 mA
Contact exposed conductive parts of the DUT which are
not connected to the protective conductor (PE) with the
probe (socket 1/2).
I-B (N/L) / Direkt 230 / GW 0,1 mA I-B (N/L) / Direkt 230 / GW 0,1 mA I-B (N/L) / Direkt 230 / GW 0,1 mA
Contact the type B application parts at the DUT with the probe
(socket 1/2).
Contact the type B application parts at the DUT with the probe
(socket 1/2).
I-PA(N/L) Direkt /230 / GW 0,1 mA I-PA(N/L) Direkt /230 / GW 0,1 mA