10.3.1 Test Sequences in Detail
(test sequences and descriptions are included on the CF card:
751NeueNorm.xml / test_sequences-0751_1_2008.pdf)
Medical devices with type CF application parts
0751 SkI CF 0751 SK II CF 0751 SKI + II CF
Visual inspection Visual inspection Visual inspection
Contact exposed conductive parts of the DUT which are con-
nected to the protective conductor (PE) with the probe
(socket 1/2).
Bitte kontaktieren Sie die Sonde (Buchse 1/2) mit
berührbaren leitfähigen Teilen des Prüflings.
Contact exposed conductive parts of the DUT which are con-
nected to the protective conductor (PE) with the probe
(socket 1/2).
R-PE / Pd 0,25 / GW 0,3 R-PE / Pd 0,25 / GW 0,3
Contact the probe (socket 1/2) with all exposed conductive
parts and the application parts of the DUT
Contact the probe (socket 1/2) with all exposed conductive parts
and the application parts of the DUT.
Attention! Mains voltage will be applied to the DUT. Attention! Mains voltage will be applied to the DUT, Attention! Mains voltage will be applied to the DUT.
I-GA (L/N) / Diff 230 / GW 0,5 mA I-GA (L/N) / Diff 230 / GW 0,5 mA
Contact exposed conductive parts of the DUT which are not con-
nected to the protective conductor (PE) with the probe
(socket 1/2).
I-B (L/N) / Direkt 230 / GW 0,1 mA I-B (L/N) / Direkt 230 / GW 0,1 mA
Contact the probe (socket 1/2) with all exposed conductive parts
and the application parts of the DUT.
Attention! Mains voltage will be applied to the DUT. Attention! Mains voltage will be applied to the DUT
.
Attention! Mains voltage will be applied to the DUT
.
I-GA (N/L) / Diff 230 / GW 0,5 mA I-GA (N/L) / Diff 230 / GW 0,5 mA
Contact exposed conductive parts of the DUT which are not
connected to the protective conductor (PE) with the probe
(socket 1/2).
I-B (N/L) / Direkt 230 / GW 0,1 mA I-B (N/L) / Direkt 230 / GW 0,1 mA
Contact the type B application parts at the DUT with the
probe (socket 1/2).
Contact the type B application parts at the DUT with
the probe (socket 1/2).
Contact the type B application parts at the DUT with the probe
(socket 1/2).
Switch the device under test on. Switch the device under test on. Switch the device under test on.
I-EPA / CF 230 / GW 0,05mA I-EPA / CF 230 / GW 0,05mA I-EPA / CF 230 / GW 0,05mA