Operation Manual

136 GMC-I Messtechnik GmbH
Appendix
10.2.2 Test sequences in detail (DIN VDE 0701-0702)
(test sequences and descriptions are included on the CF card:
0701-0702KaDrFeSt.xml / test_sequences-0701-0702_2008.pdf)
0701-0702 EDV SK I 0701-0702 EDV SK I + II 0701-0702 EDV SK II 0701-0702 aktiv SK I
Visual inspection Visual inspection Visual inspection Visual inspection
Contact exposed conductive parts of the DUT
which are connected to the protective conduc-
tor (PE) with the probe (socket 1/2).
Contact exposed conductive parts of the DUT
which are connected to the protective conduc-
tor (PE) with the probe (socket 1/2).
Contact exposed conductive parts of the DUT
which are connected to the protective conduc-
tor (PE) with the probe (socket 1/2).
R-PE / Pd 0,25 / GW 0,3 R-PE / Pd 0,25 / GW 0,3 R-PE / Pd 0,25 / GW 0,3
R-ISO / SK1 500 / GW 1M
Switch the device under test off. Switch the device under test off. Switch the device under test off. Switch the device under test off.
Attention! Mains voltage will be applied to
the DUT.
Attention! Mains voltage will be applied to
the DUT.
Attention! Mains voltage will be applied to
the DUT.
Attention! Mains voltage will be applied to
the DUT.
I-PE (L/N) / Diff 230 / GW 3,5 mA I-PE (L/N) / Diff 230 / GW 3,5 mA I-PE (L/N) / Diff 230 / GW 3,5 mA
Contact exposed conductive parts of the DUT
which are
not
connected to the protective con-
ductor (PE) with the probe
(socket 1/2).
Contact exposed conductive parts of the DUT
which are
not
connected to the protective con-
ductor (PE) with the probe
(socket 1/2).
I-B (L/N) / Direkt 230 / GW 0,5 mA I-B (L/N) / Direkt 230 / GW 0,5 mA
Attention! Mains voltage will be applied to
the DUT..
Attention! Mains voltage will be applied to
the DUT.
Attention! Mains voltage will be applied to
the DUT.
Attention! Mains voltage will be applied to
the DUT.
I-PE (N/L) / Diff 230 / GW 3,5 mA I-PE (N/L) / Diff 230 / GW 3,5 mA I-PE (N/L) / Diff 230 / GW 3,5 mA
Contact exposed conductive parts of the DUT
which are
not
connected to the protective con-
ductor (PE) with the probe
(socket 1/2).
I-B (N/L) / Direkt 230 / GW 0,5 mA I-B (N/L) / Direkt 230 / GW 0,5 mA