Operation Manual

GMC-I Messtechnik GmbH 11
Safety Warnings
Consumers with high inrush current (> 16 A) – function test
(e.g. fluorescent tubes, halogen lamps, headlights etc.) Ob-
serve the following instructions in order to prevent excessive
contact loads.
Attention!
!
The test system is protected with two 16 A FF fuses.
In order to avoid blowing the fuses, we recommend the
use of a Z850 D adapter for testing with sequence I,
making current > 16 A (Z850 B required for connection
to 32 A systems).
The measuring and test system may not be used:
If it demonstrates visible damage
With damaged connector cables, measuring cables or patient ports
If it no longer functions properly
In such cases, the instrument must be removed from operation
and secured against unintentional use.
Opening of Equipment / Repair
The equipment may be opened only by authorized service per-
sonnel to ensure the safe and correct operation of the equipment
and to keep the warranty valid.
Even original spare parts may be installed only by authorized ser-
vice personnel.
In case the equipment was opened by unauthorized personnel,
no warranty regarding personal safety, measurement accuracy,
conformity with applicable safety measures or any consequential
damage is granted by the manufacturer.
Any warranty claims will be forfeited when the warranty seal has
been damaged or removed.
Meanings of Symbols on the Instrument
300 V CAT II
Maximum permissible voltage and measuring category between
connections 1 through 4, the test socket and ground
I System with maximum 16 A nominal current
Warning regarding dangerous electrical voltage
Warning concerning a point of danger
(attention: observe documentation!)
The device may not be disposed of with the trash.
Further information regarding the WEEE mark can be
accessed on the Internet at www.gossenmetrawatt.com by
entering the search term WEEE.
Any warranty claims will be forfeited when the warranty
seal has been damaged or removed.
Data Security
Save your data to the inserted compact flash card or to a PC on a regular
basis in order to prevent any loss of data (see section 4.11.3).
We assume no liability for data loss.
We recommend the following PC program for data processing
and data administration: vFM/PS3 (transmission of measurement
data to a PC, documentation, administration, report generating
and deadline follow-up).
Condensed instructions for PS3 software for maintenance and equipment
management are provided on the CD ROM.
They include important instructions regarding the following topics:
PS3 demo with additional instructions
Special features for reading in measured values from a data file
!