Automatic Test – Administration
Test Step Specification – External Measurement
Measurement results from an external test instrument can be en-
tered for this test step. Whether or not the DUT has passed the
test or not is determined with additional entries for Min-Max limit
values and setpoint.
External measurements and visual inspections can be configured
such that voltage to the test socket is switched on during execu-
tion.
Short-form Operating Instructions
Setting Up an External Measurement
Ð Enter a new test step:
Ð Select external measurement
General Index Card
Ð Designation: Enter for purpose of identification/archiving
appears as a a headline in other windows for
improved clarity and cross-reference.
Ð Class: External measurement
Ð Description: Enter comment if required
Evaluation Index Card
The limit values must be entered here which result in failure of the
test in the event that they are exceeded (MaxLimitValue) or fallen
short of (MinLimitValue). A setpoint can be entered here as well.
See also section 6.3.6, “Evaluation of Measured Values for Indi-
vidual Test Steps”.
Ð Unit: Enter unit of measure
Unit Unit of measure for limit value setpoint or tol.
Ð Test condition: In progress
Ð Min Limit Value See entry examples in table on page 98
Ð Max Limit Value See entry examples in table on page 98
Ð Setpoint: Target value
Ð Tolerance: In progress
Note
Select a limit value which either complies with or is stricter
than the legally binding regulations.