3.2 Automatic Recognition of Mains Connection Errors
The test instrument automatically recognizes mains connection errors, 
if the conditions in the following table have been fulfilled. The user is 
informed of the type of error, and all measuring functions are disabled 
in the event of danger.
1)
In SETUP – test sequence – IT system
Attention!
!
If you discover during the protective conductor potential test 
that the mains protective conductor is charged with voltage (in ac-
cordance with the two cases mentioned first), you may no longer 
perform any measurements with your test instrument. Voltage is ac-
tually present at the exposed earthing contacts of the standard 
socket (21) as well and may be dangerous to you. Disconnect 
the test instrument from the mains immediately and see to it that 
the defect at the mains connection is repaired.
Note
+
Voltage at the electrical system’s protective conductor PE may result in 
distorted measurement values during testing for the absence of 
voltage, or during leakage voltage measurements.
4 General Notes
4.1 Online Instructions
Integrated online instructions inform the operator regarding all required 
connections, necessary work steps, operator errors, measurement results 
and more in all measuring modes. 
Information and test results appear at the dot matrix LCD in clear text.
4.1.1 Changing the User Interface Language
If you require a different language for the user interface of the test instru-
ment, you can load it into your test instrument by means of the update 
and options installation program „SECU-Up“. The program is available for 
download from our website www.gossenmetrawatt.com (Products > 
Software > Software for Testers > SECU-Up).
Upon installation on your PC and starting the program, you proceed by 
selecting the „Update“ menu and choosing a language from the following 
list: 
Deutsch, English, Français, Italiano, … .
Only one language at a time can be installed on the test instrument, the 
one previously installed is overwritten in the process.
Attention!
!
During data transmission, the test instrument and PC may not 
be disconnected from the mains power supply under any cir-
cumstances.
No other programs under WINDOWS may be activated during 
the update.
Type of Mains 
Connection Error
Message Condition Measurements
Phase conductor L 
at protective conductor 
PE to finger contact
(key)
Text appears at 
LCD
Press 
key
U > 100 V
disabled
Protective conductor PE 
and phase conductor L 
reversed and/or 
neutral conductor N 
interrupted
lamp
lights up
Voltage at PE 
>65V
impossible
(no supply power)
Contact voltage at 
protective conductor PE
to neutral conductor N
Text appears at 
LCD
U > 50 V
disabled, although 
disabling can be 
deactivated
1)
Mains voltage
too low lamp
lights up
U
L-N 
< 180 V
possible 
under certain 
circumstances
!
!