Operation Manual
GMC-I Messtechnik GmbH 63
I
AP
 Applied Parts Leakage current
Sequence
Ð Select the test: I EPA softkey.
Ð Connect the DUT to the test socket, or connect the probe.
Ð Select measuring range, test voltage Ut, test frequency ft and 
measurement duration.
Ð Start the test: Press the ONSTART key.
Ð After each reconnection to the mains, and as soon as the first 
test is started, a mains connection test is executed (see 
section 4.1.1).
Ð Contact all of the application parts, one after the other, with the 
probe.
Ð Measured values are displayed.
Ð End the test: Press the ONSTART key.
Ð Read the measured value and compare it with the table of 
allowable limit values.
Maximum Allowable Limit Values for Leakage Current in mA 
Test Standard
I
EPA
Type BF Type CF
SFC SFC
IEC 62353 
(VDE 0751-1)
Alternating 
current
5 0.05