Operation Manual
SECULIFE IF+
RUNNING A TEST
To run a flow test, the chamber must be Ready for Flow, as indicated in the base unit display (see
Messages). When the base unit is setup for Manual Reset, the key must be used to manually
reset the unit after each test. When the base unit is setup for Automatic Reset, the chamber will be
automatically reset when the water level drops below the bottom sensor.
A flow test can be started whenever the display indicates “Ready for Flow”. The flow test is started by
water being detected at the bottom sensor. Before initiating the flow on the Infusion pump under test,
ensure that the pump is programmed to deliver a sufficient amount of water to fill the chamber to the
top sensor.
NOTE: The chamber volumes of 3.5mL and 35mL are nominal; it is advised that delivery volume
should be about 4 mL and 40 mL respectively.
GMC-I Messtechnik GmbH 19