Operation Manual
SECULIFE IF+
MESSAGES
Several status messages are available to indicate the present state of the system. The following is a brief
description of the available messages:
CHAMBER NOT DETECTED – This message indicates that a Chamber has not been detected.
READY FOR FLOW– This message indicates that the Chamber is reset and waiting for fluid to be
detected at the bottom sensor. As soon as fluid is detected, the timer will begin running and the test will
begin. The test ends when the fluid reaches the top sensor or when the Reset key is depressed.
FLOW RATE TOO HIGH – This message indicates that the flow rate is higher than the resolution of the
meter.
INVALID SENSOR SEQUENCE– This message indicates that water is detected at the top sensor but
not at the bottom sensor. This could be caused by either a bad sensor or a bubble in the tube.
CHAMBER NEEDS CALIBRATION– This message indicates that a Chamber has been detected, but
the calibration data stored in the Chamber is invalid. The Chamber should be returned for re-calibration.
LOW BATTERY– This message indicates that the batteries are low and should be replaced.