Operating Manual SECULIFE IF+ Infusion Pump Analyser 3-349-623-03 1/3.
SECULIFE IF+ Contents WARNINGS, CAUTIONS, NOTICES........................................................................................................4 DESCRIPTION....................................................................................................................................... 7 OVERVIEW............................................................................................................................................ 9 KEYS....................................................
SECULIFE IF+ CAUTION - FLUID Only distilled water should be used in the chambers with the SECULIFE IF+. Do not use tap water, glucose or any other fluid; this will cause the tubing to become contaminated. CAUTION - SERVICE The SECULIFE IF+ Analyzer is intended to be serviced only by authorized service personnel. Troubleshooting and service procedures should only be performed by qualified technical personnel. CAUTION - ENVIRONMENT The SECULIFE IF+ Analyzer is intended to function between 15 and 40 °C.
SECULIFE IF+ NOTICE – SYMBOLS Symbol Description Caution (Consult Manual for Further Information) Center Negative Direct Current Per European Council Directive 2002/95/EC, do not dispose of this product as unsorted municipal waste. NOTICE – DISCLAIMER GMC-I MESSTECHNIK GMBH WILL NOT BE RESPONSIBLE FOR ANY INJURIES SUSTAINED DUE TO UNAUTHORIZED EQUIPMENT MODIFICATIONS OR APPLICATION OF EQUIPMENT OUTSIDE OF THE PUBLISHED INTENDED USE AND SPECIFICATIONS.
SECULIFE IF+ 6 GMC-I Messtechnik GmbH
SECULIFE IF+ Gossen Metrawatt SECULIFE IF+ INFUSION PUMP ANALYZER The SECULIFE IF+ is a Microprocessor based High Precision infusion Pump Analyzer (IPA). It tests the flow rate of intravenous (I.V.) infusion pumps. The flow rates are displayed in milliliters per hour. The unit can test two volumetric pumps for output flow rate simultaneously using two volumetric chambers. The following are highlights of some of the main features.
SECULIFE IF+ OPTIONAL ACCESSORIES: 120 VAC BATTERY ELIMINATOR, U.S. 220 VAC BATTERY ELIMINATOR, EUROPE SOFT CARRYING CASE OVERFLOW ASSEMBLY KIT 3.5 mL OVERFLOW ASSEMBLY KIT 35 mL AVAILABLE MODELS: The base unit plus one of the volumetric chambers are required to test infusion pumps. The part numbers are: 8 M695D SECULIFE IF+ INFUSION PUMP ANALYZER BASE UNIT Z695C SECULIFE IF1 3.
SECULIFE IF+ OVERVIEW This section looks at the layout of the SECULIFE IF+ and its Chambers and gives descriptions of the elements that are present. 8 pin Mini-Din connector for connection to Chamber 1 8 pin Mini-Din connector for connection to Chamber 2 6 Light Touch Keys for Selecting Parameters and Settings: POWER for Turning Unit On and Off UP and DOWN Arrows for Scrolling Through Selected Options. (When No Options are Selected, for Changing Display Screen.
SECULIFE IF+ 3.
SECULIFE IF+ KEYS Six tactile-touch keys are provided for system operation: – This key turns the unit off and on. The unit will return to the screen that was active when it was turned off. – In the DISPLAY MODE, these keys toggle the display through the available main screens. In the SELECT MODE, if a parameter has been highlighted, these keys with scroll through the available settings. – On the Setup screen, there are a number of parameters that may be selected and changed.
SECULIFE IF+ SCREENS MAIN SCREENS – There are four main screens: Chamber 1, Chamber 2, DUAL and Connector. The available screens can be toggled using . CHAMBER SCREENS – The Chamber screens have a large time and flow rate display, as shown below. The arrow at the top of the screen indicates the connector used for that Chamber. The level indicator at the bottom of the screen identifies the level of water in the Chamber.
SECULIFE IF+ DUAL SCREEN – The DUAL screen shows the status of both Chambers at the same time. MESSAGE DISPLAY OR DATA CONNECTOR LOCATION Chamber Not Detected Ready For Flow Flow Rate Too High Invalid Sensor Sequence Chamber Needs Calibration Low Battery Chamber 1 Chamber 2 LEVEL INDICATOR DATA UNITS See Page 16 for special graphics Seconds mL/hr CONNECTOR SCREEN – The connector screen indicates the layout of the connectors at the top of the SECULIFE IF+.
SECULIFE IF+ MESSAGES Several status messages are available to indicate the present state of the system. The following is a brief description of the available messages: CHAMBER NOT DETECTED – This message indicates that a Chamber has not been detected. READY FOR FLOW– This message indicates that the Chamber is reset and waiting for fluid to be detected at the bottom sensor. As soon as fluid is detected, the timer will begin running and the test will begin.
SECULIFE IF+ LEVEL INDICATOR A special graphic has been incorporated into the display to identify the level of water in each Chamber. The graphic is located at the lower right corner of the display for each channel. This graphic indicates that water is not present at the lower or upper sensor. This graphic indicates that fluid is present at the lower sensor. These graphics are displayed sequentially to indicate that a test is running.
SECULIFE IF+ SETUP The Setup Mode allows the user to adjust the configuration of the meter. The Setup screen can be entered using the line and key. The parameters can be changed by using key to highlight the to toggle the available options. The Setup screen can be exited using the key.
SECULIFE IF+ OPERATIONS CONNECTING CHAMBERS Chambers are connected to the base unit with a straight through 8Pin Mini Din male to 8Pin Mini Din Male cable. Chambers can be connected or disconnected at any time. Disconnecting a chamber while a test is running will cancel the test. As a chamber is connected or disconnected, the display will change to display the available chambers.
SECULIFE IF+ THEORY OF OPERATIONS Infrared sensors are used to detect the presence of water in the chamber. There are two sensors in each chamber. As water is detected at the bottom sensor, an internal timer is started. The timer runs internally in hundredths of seconds but is only displayed in tenths of seconds. The test automatically ends when water is detected at the top sensor. A buzzer will sound to indicate that the test is complete.
SECULIFE IF+ RUNNING A TEST To run a flow test, the chamber must be Ready for Flow, as indicated in the base unit display (see Messages). When the base unit is setup for Manual Reset, the key must be used to manually reset the unit after each test. When the base unit is setup for Automatic Reset, the chamber will be automatically reset when the water level drops below the bottom sensor. A flow test can be started whenever the display indicates “Ready for Flow”.
SECULIFE IF+ OPTIMIZING MEASUREMENTS To optimize the performance of the SECULIFE IF+, it is recommended that a wetting agent be used to minimize the surface tension of the test solution. The test solution should be Distilled water mixed with “MICRO-90.” “MICRO-90” can be purchased from Cole- Parmer in a 1% solution (Part Number A18100-01, www.coleparmer.com, 1-800-323-4340). Further dilute the “MICRO –90” by using one part of “MICRO-90” to 10 parts of Distilled water.
SECULIFE IF+ LIMITED WARRANTY WARRANTY: GMC-I MESSTECHNIK GMBH WARRANTS ITS NEW PRODUCTS TO BE FREE FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY ARE INTENDED. THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF SHIPMENT. EXCLUSIONS: THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR IMPLIED, INCLUDING, BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
SECULIFE IF+ SPECIFICATIONS SECULIFE IF+ 22 FLOW RANGE 3.5 mL CHAMBER 0.0 to 999.9 mL/hr FLOW RANGE 35 mL CHAMBER 15.0 to 9999.9 mL/hr FLOW RESOLUTION 0.1 mL/hr FLOW ACCURACY +/- 1% RDG, +/- LSD TIME RESOLUTION INTERNAL 0.01 Seconds DISPLAY LCD Graphical 128 X 64 Pixels SETUP MEMORY EEPROM, All Parameters MEMORY RETENTION 10 Years w/o Power OPERATING RANGE 0 to 50 Degrees C STORAGE RANGE -40 to 60 Degrees C CONSTRUCTION Enclosure - ABS Plastic Face - Lexan, Back Printed SIZE 7.
SECULIFE IF+ 3.5 mL and 35mL CHAMBERS FLOW RANGE 3.5 mL CHAMBER 0.0 to 999.9 mL/hr FLOW RANGE 35 mL CHAMBER 15.0 to 9999.9 mL/hr FLOW RESOLUTION 0.1 mL/hr FLOW ACCURACY +/- 1% RDG, +/- LSD OPERATING RANGE 0 to 50 Degrees C STORAGE RANGE -40 to 60 Degrees C CONSTRUCTION Enclosure - ABS Plastic SIZE 14.0 x 5.0 x 3.5 inches 356 x 127 x 89 mm (HxWxD) WEIGHT < 2.
Product Support If required please contact: GMC-I Messtechnik GmbH Product Support Hotline Phone +49 911 8602-0 Fax +49 911 8602-709 E-Mail support@gossenmetrawatt.com Service Center Repair and Replacement Parts Service Calibration Center * and Rental Instrument Service When you need service, please contact: GMC-I Service GmbH Service Center Thomas-Mann-Strasse 20 90471 Nürnberg • Germany Phone +49 911 817718-0 Fax +49 911 817718-253 E-Mail service@gossenmetrawatt.com www.gmci-service.