Operation Manual
21
TEMPERATURE MEASUREMENT (OPTIONAL)
RANGE
-20.0 to 100.0 °C (-4.0 to 212.0 °F)
RESOLUTION
0.1 °C (0.1 °F)
ACCURACY
± 0.5% FS
CONNECTIONS
¼” Phone Jack for use with ¼” Phone Plug terminated
temperature cables or probes.
TRANSDUCER
COMPATIBILITY
OPTION Y7
YSI 700 Transducers
OPTION RTD
100 Ω RTD
Supports 0.00385 and 0.00392 Ω/Ω/°C
temperature coefficient (alpha) sensors
ANALOG OUTPUT (OPTIONAL)
RANGE
1.0 to 4.0 VDC/FS, Selectable
ACCURACY
± 0.1% FS
RATE
Output dependent on
Digital Filter setting
CONNECTIONS
Male BNC Connector
PHYSICAL & ENVIRONMENTAL
DISPLAY
128 X 64 Pixels Non-Backlit Graphical LCD
CONSTRUCTION
ENCLOSURE
ABS Plastic
OVERLAY
Back-printed Lexan
SIZE
7.69 x 3.97 x 1.80 Inches
(195.3 x 100.8 x 45.7 mm)
WEIGHT
< 1 Lbs (0.45 kg)
OPERATING
RANGE
15 to 30 °C
(59 to 86 °F)
STORAGE
RANGE
-40 to 60 °C
(-40 to 140 °F)