Operation Manual
9
The following is a breakdown of the parameters available in the configuration of the unit
and their available options:
System Setup Configuration
Parameter
Description
Range
Display Min/Max
Selects whether the Min and Max values will be
displayed on the main screens
(except COMBINED).
yes/no
Analog Scale
Analog Output Scaling voltage. This is the
maximum analog output voltage. The output is
scaled to this voltage over the positive range of the
selected analog source.
1.0 to 4.0 Volts
Analog Source
Selects the source reading for the analog output
Pres1, Pres2, or
Temp
Contrast Adjust
Sets the contrast of the display screen.
0 to 20
Auto Off Timer
Determines the period of inactivity before the meter
is turned OFF. A timer is started when the meter is
turned ON and is reset each time a key is pressed.
When the timer reaches the value set in this
parameter, the power is automatically turned OFF.
(NOTE: Setting this parameter to 0 disables the
Auto Off timer. When running from line power, the
meter does not automatically shut off.)
0 to 30 Minutes
Battery Life
Displays current life of the battery.
At 10%, a warning screen will appear.
0 to 100%
(Read Only)
Beep Length
Sets audible beep duration.
0 to 15
Filter – Pres 1
Determines the number of samples that are
averaged in the digital filter. The software has a
Digital Filter that averages the readings to produce
a stable display.
(NOTE: Increasing this setting will cause a more
stable display. However, it will also cause a slower
response to small changes. The best setting is the
smallest number that provides a stable display.)
0 to 10 Seconds
Filter – Pres 2
Filter – Temp
RTD Type
(OPTION R1)
Sets the Temperature Coefficient (alpha) to match
that of the RTD probe.
0.00385/°C or
0.00392/°C
Software
Displays current software program.
(Read Only)