User Manual SECULIFE DPPRO DIGITAL PRESSURE METER 3-349-847-03 1/1.
GMC-I Messtechnik GmbH DP Pro TABLE OF CONTENTS WARNINGS, CAUTIONS, NOTICES ............................................................................. ii DESCRIPTION .............................................................................................................. 1 LAYOUT ......................................................................................................................... 5 SCREENS ...........................................................................................
WARNING - USERS The DP Pro is for use by skilled technical personnel only. WARNING - USE The DP Pro is intended for testing only and should never be used in diagnostics, treatment or any other capacity where it would come in contact with a patient. WARNING - CONNECTIONS All connections to patients must be removed before connecting the DUT to the DP Pro.
CAUTION - CLEANING Do not immerse. The Meter should be cleaned by wiping gently with a damp, lint-free cloth. A mild detergent can be used if desired. CAUTION - LIQUIDS Do not submerge or spill liquids on the DP Pro. Do not operate the DP Pro if it may have been exposed to fluid. CAUTION - ENVIRONMENT Exposure to environmental conditions outside the specifications can adversely affect the performance and accuracy of the DP Pro.
NOTICE – CE The DP Pro Meters bear the mark Based on the following testing standards: ELECTROMAGNETIC COMPATIBILITY DIRECTIVE EMC – Directive 89/336/EEC as amended by 92/31/EEC and 93/68/EEC EN 61326-1:1997 + A1:1998 + A2:2001 + A3:2003 “Electrical equipment for measurement, control and laboratory use – EMC requirements” This equipment has been type tested by an independent, accredited testing laboratory and compliance was demonstrated to the above standard to the extent applicable.
NOTICE – SYMBOLS Symbol Description Center Negative NOTICE – ABBREVIATIONS ANSI ASCII BCD C cmH20 ° DUT DC American National Standards Institute American Standard Code for Information Interchange Binary Coded Decimal Celsius centimeters of water degree(s) Device Under Test Direct Current Euro European F Fahrenheit FS inHg inH20 kg 2 kg/cm Full Scale inches of mercury inches of water kilogram(s) kilogram(s) per centimeter squared kHz kilohertz kPa kilopascal(s) Max Maximum µA microampere(s)
NOTICE – DISCLAIMER GMC-I Messtechnik GmbH WILL NOT BE RESPONSIBLE FOR ANY INJURIES SUSTAINED DUE TO UNAUTHORIZED EQUIPMENT MODIFICATIONS OR APPLICATION OF EQUIPMENT OUTSIDE OF THE PUBLISHED INTENDED USE AND SPECIFICATIONS. NOTICE – DISCLAIMER GMC-I Messtechnik GmbH RESERVES THE RIGHT TO MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT.
GMC-I Messtechnik GmbH DP Pro DIGITAL PRESSURE METERS The Model DP pro is a family of microprocessor-based, high-precision Pressure Meters, which are intended for use in the evaluation and servicing of a wide variety of medical, commercial and industrial applications. These meters measure compatible gas and liquid pressures in various engineering units.
TEMPERATURE OPTION ADDS: OPTION Y7 - YSI 700 TEMPERATURE PROBE INTERFACE OPTION R1 - 100 Ω RTD TEMPERATURE PROBE INTERFACE -20.0 TO 100.0 °C / -4.0 TO 212.0 °F TEMPERATURE RANGE ± 0.
MODEL INFORMATION SPECIAL COUPLER CONFIGURATIONS Option Pressure 1 Coupler (Right Port) Pressure 2 Coupler (Left Port) S01 S02 S03 S04 S05 S06 S07 S08 S09 S10 S11 S12 Female Quick-Disconnect Female Quick-Disconnect Female Quick-Disconnect Female Quick-Disconnect Male Quick-Disconnect Male Quick-Disconnect Male Quick-Disconnect Male Quick-Disconnect Male Luer Lock Male Luer Lock Male Luer Lock Male Luer Lock Female Quick-Disconnect Male Quick-Disconnect Male Luer Lock Not Applicable Female Quick-Discon
PRESSURE RANGES BY SENSOR RANGE PRESSURE UNITS PSI mmHg @ 0° C mmHg @ 20° C inHg @ 0° C inHg @ 20° C cmH2O @ 20° C inH2O @ 4° C inH2O @ 20° C inH2O @ 60° F kg/cm 2 kPa mBar Bar PRESSURE SENSOR RANGE 100 PSI 75 PSI 10 PSI 5 PSI 0.3 PSI -13.500 to 100.000 -698.2 to 5171.5 -700.6 to 5190.3 -27.486 to 203.602 -27.586 to 204.342 -951.8 to 7043.2 -373.6 to 2768.1 -374.3 to 2772.9 -374.1 to 2770.8 -.9491 to 7.0306 -93.08 to 689.48 -930.8 to 6894.8 -.9308 to 6.8948 -13.500 to 75.000 -698.2 to 3878.6 -700.
LAYOUT This section looks at the layout of the DP Pro and gives descriptions of the elements that are present. Pressure Port #2: Male Luer coupler pictured Analog Output: BNC (Optional) Temperature: ¼” Phone Jack (Optional) NOTE: Coupler type depends on sensor and options, see Model Information for details. Pressure Port #1: Female QuickDisconnect pictured NOTE: Coupler type depends on sensor and options, see Model Information for details.
SCREENS MAIN SCREENS – There can be up to four main screens, depending on the model. They are PRESSURE 1, PRESSURE 2, TEMPERATURE, COMBINED and INPUTS. The available screens can be toggled through using .
Input Identification Screen Note: Sensor limits are displayed based on selected range. PRESSURE SCALE – The pressure scale is indicated by the units displayed under the reading. The scale can be changed by using to highlight the unit line and to toggle between available pressure units as listed below.
TEMPERATURE SCALE – The temperature scale is indicated by the units displayed under the reading. The scale can be changed by using to highlight the unit line and to toggle the temperature units between Degrees Celsius (°C) and Fahrenheit (°F). NOTE: If the measured temperature is outside of the range of the instrument, an OVER RANGE or UNDER RANGE message box will be displayed. For models with the YSI option, the NO PROBE message box will be displayed when the unit detects an open connection.
The following is a breakdown of the parameters available in the configuration of the unit and their available options: System Setup Configuration Parameter Display Min/Max Analog Scale Description Selects whether the Min and Max values will be displayed on the main screens (except COMBINED). Analog Output Scaling voltage. This is the maximum analog output voltage. The output is scaled to this voltage over the positive range of the selected analog source.
ZEROING PRESSURE SCALES – When there is no pressure applied to either port, the display should read “0.” It may be necessary to zero the pressure scales to remove any errors due to ambient conditions. This is done by pressing the zeroing instructions are displayed, then pressing key until the simultaneously to begin the process. The “ZEROING…” message will flash while the scale is being zeroed. When the zeroing instructions are displayed again, the process is complete.
KEYS Six tactile-touch keys are provided for system operation: – This key turns the unit off and on. The unit will return to the main screen that was active when it was turned off. – In the DISPLAY MODE, these keys toggle the display through the available main screens. In the SELECT MODE, if a parameter has been highlighted, these keys will scroll through the available settings. – On any screen, there are a number of parameters that may be selected and changed.
– This key toggles the unit into and out of the Setup Mode. Depressing this key will enter the Setup screen where the configuration can be viewed and adjusted. Depressing the key again will exit the Setup Mode and return to the previously viewed main screen. This will also save any changes to the internal EEPROM memory so they will be retained even with the power turned off or battery removed.
OPTIONS ANALOG OUTPUT – The unit may be ordered with a DC Analog Output Option. This option provides a filtered analog output that is representative of the displayed pressure or temperature, and is provided via a BNC connector on the top of the unit. The source parameter for the analog output is selectable in the Setup Menu between Pressure (Pres) or Temperature (Temp). The output is scaled to match the 0 to FS range of the selected source parameter over a variable internally generated reference voltage.
COMMUNICATIONS Since the meter does not handle a great deal of data, the RS-232 communications link has been optimized to allow the user, through very simple instructions, to control and request data from the meter. Refer to Specifications section for RS-232 Settings (Baud, etc). Data transmitted/received is in standard ASCII format, and all numerical values are in BCD format. All commands sent to the unit should be terminated with a “Carriage Return” character ( or in hexadecimal, 0x0D).
Examples: Data Sent W064 Data Returned W064 W0600004 W0600004 W05100 ?? U - UPLOAD Meaning Echo of Command Sent (Set Pressure units to “inH2O”) Echo of Command Sent (Set Pressure units to “inH2O”) W05100 Echo of Command Sent Invalid Command Response (Location 05 is Read Only) The UPLOAD command allows the user to read all of the selected device data from locations 1 through 16 with a single command.
X - CANCEL The CANCEL command is simply a way to re-establish proper control, should a communications error occur or an incorrect command be transmitted. For the most part, an incorrect command will simply be ignored and the meter will return to listening for future commands. However, a prior command may be cancelled midstream by transmitting the CANCEL command anytime. Usage: X This command does not require a carriage return, nor will it acknowledge with a carriage return.
DATA LOCATIONS LOCATION ACCESS DESCRIPTION RANGE 01 02 03 READ READ/WRITE READ/WRITE BATTERY LIFE REMAINING CONTRAST AUTO POWER OFF 0 to 100% 0 to 20 0 to 30 (seconds) 04 READ MODEL RESERVED 05 READ PRESSURE 1 TYPE 06 READ/WRITE PRESSURE 1 UNITS 07 08 09 10 READ/WRITE READ READ/WRITE READ/WRITE PRESSURE 1 FILTER PRESSURE 1 PRESSURE 1 MAX PRESSURE 1 MIN 11 READ PRESSURE 2 TYPE 12 READ/WRITE PRESSURE 2 UNITS 13 14 15 16 READ/WRITE READ READ/WRITE READ/WRITE PRESSURE 2 FILTER PRESS
DATA LOCATIONS (Temp option only) LOCATION ACCESS DESCRIPTION 17 READ TEMPERATURE SENSOR TYPE 18 READ/WRITE TEMPERATURE UNITS 19 20 21 22 READ/WRITE READ READ/WRITE READ/WRITE TEMPERATURE FILTER TEMPERATURE TEMPERATURE MAX TEMPERATURE MIN RANGE 0 1 2 0 1 Not Applicable YSI 700 RTD 100 °C °F 0-60 See Note 2 See Note 2, 3 See Note 2, 3 NOTES 1. Pressure readings are returned in the currently set Pressure Units parameter in Location 6.
MANUAL REVISIONS Revision # Program # Revisions Made Rev 01 Rev 02 Rev 03 Rev 04 Rev 05 DT7321CA DT7321CB DT7321CB DT7321CG DT7321CH Rev 06 DT7321CI Origination Min/Max made standard, Quick-Disconnect added MC and FC option for Quick-Disconnect added Misc. Updates Format Updated, Specifications Updated, Misc.
SPECIFICATIONS PRESSURE MEASUREMENT (GAUGE) RANGE RESOLUTION MAXIMUM OVERPRESSURE 100 PSI SENSOR -13.5 TO 100.0 PSI 75 PSI SENSOR -13.5 TO 75.0 PSI 10 PSI SENSOR -10.0 TO 10.0 PSI 5 PSI SENSOR -5.0 TO 5.0 PSI 0.3 PSI SENSOR -0.3 TO 0.3 PSI 100 PSI SENSOR 0.001 PSI 75 PSI SENSOR 0.001 PSI 10 PSI SENSOR 0.0001 PSI 5 PSI SENSOR 0.0001 PSI 0.3 PSI SENSOR 0.00001 PSI 100 PSI SENSOR 200 PSI 75 PSI SENSOR 200 PSI 10 PSI SENSOR 45 PSI 5 PSI SENSOR 15 PSI 0.
TEMPERATURE MEASUREMENT (OPTIONAL) RANGE -20.0 to 100.0 °C (-4.0 to 212.0 °F) RESOLUTION 0.1 °C (0.1 °F) ACCURACY ± 0.5% FS CONNECTIONS ¼” Phone Jack for use with ¼” Phone Plug terminated temperature cables or probes. OPTION Y7 TRANSDUCER COMPATIBILITY YSI 700 Transducers 100 Ω RTD OPTION RTD Supports 0.00385 and 0.00392 Ω/Ω/°C temperature coefficient (alpha) sensors ANALOG OUTPUT (OPTIONAL) RANGE 1.0 to 4.0 VDC/FS, Selectable ACCURACY ± 0.
ELECTRICAL & MISC.
NOTES 23