Operation Manual

GMC-I Messtechnik GmbH 83
SECULIFE DF
BASE
ENERGY OUTPUT MEASUREMENT
OTHER
OSCILLOSCOPE OUTPUT
HIGH MEASURE RANGE
1000:1 amplitude-attenuated
LOW MEASURE RANGE
200:1 amplitude-attenuated
WAVEFORM PLAYBACK
OUTPUT
LEAD I & PLATES
SCREEN
200:1 Time Base Expansion
SYNC TIME MEASUREMENTS
TIMING WINDOW
Starts at peak of
each R-wave
TEST WAVEFORMS
All waveform simulations available
DELAY TIME ACCURACY
+/- 1 ms
CHARGE TIME MEASUREMENT
From 0.1 to 99.9 sec
ECG NSR
RATE
30,40,45,60,80,90,100,120,140,160,
180,200,220,240,260,280,300 BPM
ACCURACY
+/- 1%
AMPLITUDE
0.5,1.0,1.5,2.0 mV (Lead II)
ACCURACY
+/- 2% @ Lead II
HIGH LEVEL
200 times Amplitude
ACCURACY
+/- 5%
QRS DURATION
80ms
ECG PERFORMANCE
SINE WAVE
0.1,0.2,0.5,5,10,40,50,60,100 Hz
SQUARE WAVE
0.125, 2.000 Hz
TRIANGLE WAVE
2.000, 2.500 Hz
PULSE WAVE
30,60,120 BPM; 60 ms width
AMPLITUDE
0.5,1.0,1.5,2.0 mV (Lead II)
RATE ACCURACY
+/- 1%
AMPLITUDE ACCURACY
+/- 2% @ Lead II
ECG GENERAL
LEAD TO LEAD
IMPEDANCE
(RL, LL, RA, LA)
1000 ohm
LEAD TO LEAD
IMPEDANCE
(V1-V6)
1000 ohm