ECG CONFIGURATION SECTION
This section configures the ECG Performance sequence when the ECG
Performance Test option is set to YES for a Defibrillator Test.
The ECG sequence consists of up to 10 steps. Each step consists of a
selected waveform group (Disabled, NSR, AED, Arrhythmia or
Performance), output waveform and output amplitude.
There is only one ECG sequence for all 50 Auto Sequence tests.
To use less than 10 steps, set the ECG output group of the next step
after the last to “Disabled.”