Operation Manual

76 GMC-I Messtechnik GmbH
SECULIFE DF
BASE
SEQUENCE CONFIGURATION
The sequence configuration window displays all of the configuration settings for each Auto Sequence. This
sample screen shows a defib test configuration.
Sequence Name – This name can be any string of up to 20
standard ASCII characters. NOTE: Not all ASCII characters are
valid and will be ignored.
Copy Sequence Button – This button opens the Copy
Sequence Screen that allows the user to quickly configure
similar test sequences.
Test Mode Selections – These selections allow each test to be
configured as a Defibrillator Only, a Pacer Only or a Defib &
Pacer test.
Configure Test Buttons – These buttons are used to alternate
between defib and pacer test configuration windows.
Defibrillator Test Details – Each of the potential tests and test
details for the Defibrillator are displayed for configuration. For
ease of programming, individual steps can be deleted or added
and individual tests can be included or not included.
Insert and Delete Steps Buttons – These buttons will open the
Insert Steps Screen or the Delete Steps Screen.