Operation Manual
32 GMC-I Messtechnik GmbH
SECULIFE DF
BASE
The following is a brief description of how the SECULIFE DF
BASE
simulates the available Arrhythmias:
Abbreviation Arrhythmia Description
Irregular waveform with no real
Vent Fib – Fine
Ventricular Fibrillation
P-wave or clear R-R interval and a
low signal level
(Continuous)
Atrial Fib
A
bsence of P-wave, irre
g
ular P-R
Atrial Fibrillation
interval rate and a high level signal
(Continuous)
80 BPM with increasin
g
2
nd
Deg Heart
Second Degree Heart Block
P-R interval for four beats
Block
(160
,
220
,
400
,
470
ms
)
followed by a P wave without a QRS
(Continuous)
Rt Bundle
80 BPM with Normal P-wave and P-
Right Bundle Branch Block R interval but wider QRS complexes
Branch Block
(Continuous)
NSR of 80 BPM with Periodic
PAC
Premature Atrial Contraction
Abnormal 25% early P waves
(PAC, 7 NSR)
(Continuous)
NSR of 80 BPM with periodic left
PVC Early
Early Type 1
focus premature ventricular beats
with 33% premature timing
Premature Ventricular Contraction
(PVC Type 1, 9 NSR)
(Continuous)
NSR of 80 BPM with periodic left
PVC Std
Standard Type 1
focus premature ventricular beats
with 20% premature timing
Premature Ventricular Contraction
(PVC Type 1, 9 NSR)
(Continuous)
NSR of 80 BPM with periodic left
focus premature ventricular beats
PVC R on T
R on T Type 1 with 65% premature timing, placing R
Premature Ventricular Contraction on the previous T
(PVC Type 1, 9 NSR)
(Continuous)
Multifocal
NSR of 80 BPM with T
y
pe 1 and
Multifocal Type 2 PVCs (PVC Type 1, 2 NSR,
PVCS
Premature Ventricular Contraction PVC Type 2, 2 NSR)
(Continuous)