Operation Manual

18 GMC-I Messtechnik GmbH
SECULIFE DF
BASE
General Operation
The unit is controlled by 7 light touch keys. They allow the user to move around within the displayed
parameters, select the desired options, choose a specific category and control the setup for the unit.
When a key is depressed there is an audio click when it is accepted, or a razz tone if the key is invalid.
A large LCD graphics display with backlight provides the user with information about the current status of
the device configuration options, test results and more. The display identifies the function of each key on a
dynamic basis. As the operation mode changes, the key functions change to suit the operating mode.
Range Key
The key scrolls through the ranges of the SECULIFE DF
BASE
analyzer. Depressing the key will
allow the user to select between High Defibrillator Range (1000J max), Low Defibrillator Range (50J
max) and Pacemaker Range. The default mode on power up is High Defibrillator Range.
Backlight Key
The Graphic LCD display may be viewed with or without the backlight. Depressing any key will activate
the backlight. However, since the backlight will drain the battery if left on, it will automatically shut off
after a user programmable delay when running on battery power.
The key is provided to toggle the backlight on or off at any time.