Operation Manual

12 GMC-I Messtechnik GmbH
SECULIFE DF
BASE
ENERGY OUTPUT MEASUREMENT OTHER
OSCILLOSCOPE OUTPUT
HIGH MEASUREMENT RANGE: 1000:1 amplitude-attenuated
LOW MEASUREMENT RANGE: 200:1 amplitude-attenuated
WAVEFORM PLAYBACK
OUTPUT – LEAD 1 & PLATES
GRAPHICS SCREEN
200:1 Time Base Expansion
SYNC TIME MEASUREMENTS
TIMING WINDOW: Starts at peak of each R-wave
TEST WAVEFORMS: All waveform simulations available
CHARGE TIME MEASUREMENT
From 0 .1 to 99.9 sec
ECG FUNCTIONS
The unit can produce a wide variety of ECG simulations. The user simply selects the parameters that
match the desired output.
RATE: 30,40,45,60,80,90,100,120,140,160,180,200,220,240,260,280,300 BPM
AMPLITUDE: 0.50,1.0,1.5,2.0 mV (Lead II)
ECG-PERFORMANCE FUNCTIONS
The unit can generate Sine, Square, Triangular, and Pulse waveforms with adjustable amplitudes for
performance testing.
SINE: 0.1,0.2,0.5,5,10,40,50,60,100 Hz
SQUARE: 0.125,2 Hz
TRIANGLE: 2,2.5 Hz
PULSE: 30,60,120 BPM; 60 ms WIDTH
AMPLITUDE: 0.5,1.0,1.5,2.0 mV (Lead II)