Operation Manual
GMC-I Messtechnik GmbH 11
SECULIFE DF
BASE
ENERGY OUTPUT MEASUREMENT GENERAL
The unit measures the energy in the output pulse of both monophasic and biphasic defibrillators.
• PULSE TYPE: Monophasic or Biphasic
• LOAD RESISTANCE: 50 ohm +/- 1%, non-inductive (<1 µH)
• DISPLAY RESOLUTION: 0.1 Joules
• MEASUREMENT TIME WINDOW: 100 ms
• ABSOLUTE MAX PEAK VOLTAGE: 6000 Volts
• CARDIOVERSION DELAY: 0 to 6000 ms
• CARDIOVERSION RESOLUTION: 0.1 ms
ENERGY OUTPUT MEASUREMENT HIGH RANGE
The high range allows for a large pulse with high voltage and current.
• VOLTAGE: <5000 Volts
• MAX CURRENT: 100 Amps
• MAX ENERGY: 1000 Joules
• TRIGGER LEVEL: 100 Volts
• PLAYBACK AMPLITUDE: 1 mV / 1000 V Lead I
• TEST PULSE: 125 Joules +/- 20%
ENERGY OUTPUT MEASUREMENT LOW RANGE
The low range allows greater resolution on smaller pulses.
• VOLTAGE: <1000 Volts
• MAX CURRENT: 20 Amps
• MAX ENERGY: 50 Joules
• TRIGGER LEVEL: 20 Volts
• PLAYBACK AMPLITUDE: 1 mV / 1000 V Lead I
• TEST PULSE: 5 Joules +/- 20%