Operating O Ma anual SECU S ULIFE E DFBASE B DEFIBRIL D LLATOR ANALYZE A ER 3-349-80 04-03 1/6.
SECULIFE DFBASE Contents WARNINGS, CAUTIONS, NOTICES....................................................................................................... 3 DESCRIPTION ........................................................................................................................................ 9 OVERVIEW ............................................................................................................................................. 16 DEFIBRILLATOR ANALYZER .........................
SECULIFE DFBASE WARNING - USERS The SECULIFE DFBASE analyzer is for use by skilled technical personnel only. WARNING - USE The SECULIFE DFBASE Analyzer is intended for testing only and it should never be used in diagnostics, treatment or any other capacity where it would come in contact with a patient. WARNING - MODIFICATIONS The SECULIFE DFBASE Analyzer is intended for use within the published specifications.
SECULIFE DFBASE WARNING - LIQUIDS Do not submerge or spill liquids on the Analyzer. Do not operate the Analyzer if internal components not intended for use with fluids may have been exposed to fluid, as the internal leakage may have caused corrosion and be a potential hazard. CAUTION - SERVICE The SECULIFE DFBASE Analyzer is intended to be serviced only by authorized service personnel. Troubleshooting and service procedures should only be performed by qualified technical personnel.
SECULIFE DFBASE NOTICE – IN NDICA ATIONS S FOR U USE The SECULIF S FE DFBASE Analyze er is used d to determin d ne that de efibrillato ors and transcutaneous pacem makers arre perforrming with hin their p performa ance spe ecificatio ons throug gh the me easurem ment of en nergy ou utput.
SECULIFE DFBASE NOTICE – ABBREVIATIONS 6 A, Amps Amperes BPM Beats Per Minute c C ° centi- (10 ) Celsius degree dt Delta Time, Change in Time DUT Device Under Test E Energy ECG Electrocardiogram Euro European Hz kg lbs hertz ) kilograms pounds µ micro- (10 ) µA microampere µH µV microhertz microvolt µsec microsecond m mA milli- (10 )milliampere mm ms, mS, msec mV Ω millimeter millivolts ohm P Power ppm pulse per minute R Resistance, ohms Sec, S seconds US United St
SECULIFE DFBASE GMC-I Messtechnik GmbH 7
SECULIFE DFBASE NOTICE – DISCLAIMER USER ASSUMES FULL RESPONSIBILITY FOR UNAUTHORIZED EQUIPMENT MODIFICATIONS OR APPLICATION OF EQUIPMENT OUTSIDE OF THE PUBLISHED INTENDED USE AND SPECIFICATIONS. SUCH MODIFICATIONS OR APPLICATIONS MAY RESULT IN EQUIPMENT DAMAGE OR PERSONAL INJURY. NOTICE – DISCLAIMER GMC-I MESSTECHNIK GMBH RESERVES THE RIGHT TO MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT.
SECULIFE DFBASE DEFIBRILLATOR ANALYZER SECULIFE DFBASE is a microprocessor-based instrument that is used in the testing of defibrillators. It measures the energy output and provides information about the defibrillation pulse. It is used on manual, semi-automatic and automatic defibrillators with monophasic or biphasic outputs. SECULIFE DFBASE additionally provides a Transcutaneous Pacemaker analysis function.
SECULIFE DFBASE The following are highlights of some of the main features: GENERAL • SIMPLE TO OPERATE • GRAPHICS DISPLAY WITH SIMULTANEOUS DETAILED STATUS OF PARAMETERS AND SCROLLING CONTROL OF OPTIONS 10 • ON SCREEN VIEWING OF DEFIBRILLATOR AND PACEMAKER WAVEFORMS • DROP DOWN CHOICE SCREENS LIST ALL OPTIONS FOR PARAMETERS • MONOPHASIC AND BIPHASIC COMPATIBLE • 5000 V, 1000 JOULE CAPACITY • HIGH AND LOW RANGES • CARDIOVERSION DELAY MEASUREMENT • CHARGE TIME MEASUREMENT • WAVEFORM STORA
SECULIFE DFBASE ENERGY OUTPUT MEASUREMENT GENERAL The unit measures the energy in the output pulse of both monophasic and biphasic defibrillators. • PULSE TYPE: Monophasic or Biphasic • LOAD RESISTANCE: 50 ohm +/- 1%, non-inductive (<1 µH) • DISPLAY RESOLUTION: 0.1 Joules • MEASUREMENT TIME WINDOW: 100 ms • ABSOLUTE MAX PEAK VOLTAGE: 6000 Volts • CARDIOVERSION DELAY: 0 to 6000 ms • CARDIOVERSION RESOLUTION: 0.
SECULIFE DFBASE ENERGY OUTPUT MEASUREMENT OTHER OSCILLOSCOPE OUTPUT • HIGH MEASUREMENT RANGE: 1000:1 amplitude-attenuated • LOW MEASUREMENT RANGE: 200:1 amplitude-attenuated WAVEFORM PLAYBACK • OUTPUT – LEAD 1 & PLATES • GRAPHICS SCREEN • 200:1 Time Base Expansion SYNC TIME MEASUREMENTS • TIMING WINDOW: Starts at peak of each R-wave • TEST WAVEFORMS: All waveform simulations available CHARGE TIME MEASUREMENT • From 0 .1 to 99.
SECULIFE DFBASE ARRHYTHMIA FUNCTIONS The unit can simulate 12 different arrhythmias.
SECULIFE DFBASE ACCESSORIES BC20 - 40032 INTERNAL PADDLE ADAPTERS (2 adapters) BC20 - 21103 BATTERY ELIMINATOR (120 VAC) (US Version) BC20 - 21101 BATTERY ELIMINATOR (220 VAC) (Euro Version) BC20 - 00427 PLASTIC ELECTRODE PLATES (2 plates) OPTIONAL ACCESSORIES BC20 - 30108 MEDIUM SOFT SIDED CARRYING CASE BC20 41341 COMMUNICATION CABLE (DB 9 M to DB 9 F) BC20 - 00420 PHYSIO-CONTROL DEFIB / PACE TEST CABLE BC20 - 00421 MARQUETTE DEFIB / PACE TEST CABLE BC20 00423 ZOLL DEFIB/PACE TEST CABLE BC20 - 004
SECULIFE DFBASE This page intentionally left blank.
SECULIFE DFBASE O OVERV VIEW This secttion looks at the layoutt of SECULIIFE DFBASE and a gives descriptions of the elem ments that are present. LCD G Graphical Dis splay: Showss Parameters s for Test Data a and Waveforms 5 Light Touch Keys for Dyna amic fu unctions: 10 Unive ersal Patient L Lead Connecto ors: RA R Th hese keys are e la abelled in the bo ottom portion of th he screen and d ch hange functio on ba ased on op perating mode.
SECULIFE DFBASE This section looks at the layout of the back and gives descriptions of the elements that are present. Oscilloscope output BNC connector for easy viewing of input waveforms Power Switch AC Adaptor/ Battery Eliminator 2.
SECULIFE DFBASE General Operation The unit is controlled by 7 light touch keys. They allow the user to move around within the displayed parameters, select the desired options, choose a specific category and control the setup for the unit. When a key is depressed there is an audio click when it is accepted, or a razz tone if the key is invalid. A large LCD graphics display with backlight provides the user with information about the current status of the device configuration options, test results and more.
SECULIFE DFBASE Function n Keys There are e five keys th hat are used d to provide general op perational coontrol. The functions off the keys vary depen nding on the e current sccreen. The section s of th he screen juust above th he key indicatess its current meaning. NOTE: O Only function ns that are available to o the user will w be visible e at any givven time. Sample Function Key y Labeling ECG Wa aveforms The micrroprocessorr has stored d in its mem mory all of th he digitalized d waveform ms.
SECULIFE DFBASE Universal Patient Lead Connectors The 10 Universal Patient Lead Connectors allow for 12 lead ECG simulations. AHA and IEC colorcoded labels are located on the face of the unit to aid in connecting the corresponding U.S. and International Patient Leads. AHA Label IEC Label RA R Right Arm LA L Left Arm RL N Right Leg (reference or ground) LL F Left Leg V1 V2 V3 V4 V5 V6 20 C1 C2 C3 C4 C5 C6 Description V Leads (V1-V6) (U.S.
SECULIFE DFBASE High Level Output (+) A high level ECG output signal (200 X Amplitude Setting) is available on the RCA jack located on the rear of the unit. Serial Port A female 9-pin D-Sub connector is provided for the connection of the unit to a PC or laptop serial port (e.g. Com 1). This link is then used for either remote control or flash downloading of software upgrades. Parallel Port A female 25-pin D-Sub connector is provided for the connection of a printer via a Centronics parallel interface.
SECULIFE DFBASE Power Switch A rocker switch is provided on the rear of the unit to turn the power on and off. Power Supply The unit utilizes two 9 Volt Alkaline Batteries in the bottom battery compartments. When the unit detects a LOW BATTERY condition (10% Battery Life), a warning window will appear once per minute to alert the user. Battery Eliminator The unit has a 2.1 mm micro jack for connecting a 10-Volt AC battery eliminator. The adapter will power the unit, but will not charge the battery.
SECULIFE DFBASE DEFIBR D ILLATOR ANAL LYZER MAIN SCREEN S N When the SECULIF FE DFBASE is first pow wered up, the t Defibrillator Analyz yzer MAIN SCREEN will w be displayed d. This scre een shows the currentt CONFIGU URATION, the t TEST R RESULTS and a the ava ailable FUNCTIO ON KEYS. All defibrillator tests a are run from m the MAIN N SCREEN N. When the e unit deteccts an input of g greater than n 100 Volts on the Deffibrillator Pllates (20 Vo olts in Low Range), it will automa atically begin a test.
SECULIFE DFBASE The key may m be use ed to toggle e the unit to t the Low Range Deefibrillator mode. m This mode allows for waveform ms up to 50 Joules J to be e analyzed.. The Defibrrillator Analyyzer works the same in n both ranges. T The lower ra ange simply y provides ffor a higher resolution for f pulses w with smallerr amplitudess.
SECULIFE DFBASE CONFIG C URATIO ON The CONFIGUR C RATION secction of the MAIN M SCRE EEN display ys the curre ent setup off the unit. RANG GE The first f line dissplays the range r value e for the pu ulse. It may be either 1000 1 Jouless or 50 Jou ules max. This setting s mayy be change ed using the e kkey. STAT TUS This line provide es informatio on about the e current sttatus of the analyzer. ECG This line displayss the selecttion that is active a on th he ECG term minals.
SECULIFE DFBASE TEST RES SULTS The TES ST RESULT TS section of o the MAIN SCREEN displays d the e results of tthe last puls se. It will continue to be displa ayed until th he power iss turned off, another tes st is run or tthe range is s changed. NOTE: T The unit auto omatically starts s a test when it see es a voltage e greater thhan 100 Volts on the Defibrillator Plates (20 Volts in Low Range e). NOTE: T Test results are immediately sent tto the printe er port as so oon as the ddata is available.
SECULIFE DFBASE DELAY This line normally displays the delay from the peak of the R wave until the start of the Defib Energy pulse. The line is replaced by the CHARGE TIME if this test has been run (see START CHARGE TIMER SCREEN for more information). CHG TIME This line displays if the Charge Timer has been run. It shows the time required to charge the Device Under Test (DUT). This test is started with the GMC-I Messtechnik GmbH key.
SECULIFE DFBASE FU UNCTIO ON KEYS S The FUN NCTION KE EYS section n of the MA IN SCREEN N displays the current functions of o the keys found below th he display. These key ys allow forr navigation n to supporrting screenns and inittiation of sp pecific features. Priimary Function Keys Seecondary Fun nction Keys ECG WA AVEFORMS S This key enters the ECG WAVE EFORMS sccreen wherre all ECG parameters p are set.
SECULIFE DFBASE AUTO SEQUENCES This key brings up the AUTO SEQUENCE MENU, which is used to view or run the Auto Sequences stored in the unit. SELF TEST WAVEFORM This key sends an internal test pulse to the unit, allowing for the display of the results to give an indication that the system is working properly. SECULIFE DFBASE SETUP This key brings up the SYSTEM CONFIGURATION SCREEN, which allows for adjusting the various system configuration parameters.
SECULIFE E DFBASE EC CG WAV VEFORM MS SCREEN The SEC CULIFE DFBASE ECG ou utput can be e connected d in 3, 5 or 12 lead connfigurations s. B Pressing the key fro om the MAIIN SCREEN N will allow the user to configure the waveform that is ussed for the ECG E outputt.
SECULIFE DFBASE The ECG Group, Waveform and Amplitude can be selected using parameter and using to highlight the to open a drop down menu of all the options for the highlighted parameter. Use to scroll to the desired option. Then is used to accept the new setting. The key can be used to return to the ECG waveform configuration screen without making a new selection. The key is used to return to the MAIN SCREEN.
SECULIFE DFBASE The following is a brief description of how the SECULIFE DFBASE simulates the available Arrhythmias: Abbreviation Arrhythmia Vent Fib – Fine Ventricular Fibrillation Atrial Fib Atrial Fibrillation 2 nd Deg Heart Block Second Degree Heart Block Rt Bundle Branch Block Right Bundle Branch Block PAC Premature Atrial Contraction PVC Early Early Type 1 Premature Ventricular Contraction PVC Std Standard Type 1 Premature Ventricular Contraction PVC R on T R on T Type 1 Premature
SECULIFE DFBASE Abbreviation Arrhythmia Bigeminy Bigeminal Rhythm Run of 5 PVCs Run of 5 Premature Ventricular Contractions Vent Tach Ventricular Tachycardia GMC-I Messtechnik GmbH Description NSR of 80 BPM with every other beat a Type 1 PVC (Continuous) NSR of 80 BPM with periodic group of 5 Type 1 PVCs (5 PVC Type 1, 36 NSR) (Continuous) 160 BPM, No P-wave, Beats similar to Type 1 PVC (Continuous) 33
SECULIFE DFBASE PLAYB BACK LA AST PUL LSE SCR REEN The SEC CULIFE DFBASE can display a gra aphical reprresentation of the last pulse. This s screen may m be key from the Defibriillator Anal yzer MAIN N SCREEN. The accessed d by presssing the playbackk allows the e user to vie ew the Defiibrillator pulse in a tim me-expandeed form. Samples are stored s internallyy at 0.1 ms intervals. The PLAYBA ACK LAST PULSE SCREEN show ws these sa amples expa anded by a time e factor of 200.
SECULIFE DFBASE The key starts a playback of only the first 20 ms of the waveform. The key starts a playback of the entire 100 ms of the waveform. At any time, the key or key can be depressed to return to the MAIN SCREEN.
SECULIFE DFBASE START T CHAR RGE TIMER SCR REEN A special timer has been b incorpo orated into th he SECULIF FE DFBASE to o analyze thhe charging circuit c of the e Device Under Test (D DUT). The START S CHA ARGE TIMER R SCREEN can be acceessed by pre essing the keyy from the MA AIN SCREE EN. To synch hronize the charge timeer with the de efibrillator charge time, a Pre-W Warning Co ountdown pe eriod is starrted. When the timer reeaches zero o, the defibrillattor charge should s be in nitiated.
SECULIFE DF D BASE The DUT T should be e discharge ed as soon as it becom mes charge ed. When thhe DUT is discharged d, the timer will automatica ally stop. Th he display w will show the e results of the Defibril lator pulse analysis ass well me required d to charge the t DUT: as the tim Charge C Timerr Results R At any tim me, the key can be dep pressed to end e the time er and returrn to the MAIN SC CREEN.
SECULIFE DFBASE PRIN NT HEAD DER The SEC CULIFE DFBASE provide es a heade er for record ding test data as well aas the resu ults of each pulse B that is disscharged in nto the unit. Test resultts are imme ediately sent to the prinnter port as soon as the e data is availab ble. The hea ader is sentt by pressin ng the om the MAI N SCREEN N. key fro The statu us line of the configuration section n will indicatte that the header h has been sent to t the printe er.
SECULIFE DFBASE The following is the print header and sample data that are used for the Defibrillator Analyzer mode. Gossen Metrawatt SECULIFE DFBASE Defibrillator Analyzer SECULIFE DFBASE Serial Number: Dut Manufacturer: Dut Model: Dut Serial Number: Technician: Location: Date: Comments: SECULIFE DFBASE Measurements: NOTE: Printing the header also resets the test number printed on the data sheet. NOTE: In the test results, the user must manually write the power setting of the DUT.
SECULIFE DFBASE SELF TEST WAVEFO W ORM The SEC CULIFE DFBASE has built in test wa aveforms that will give an indicatioon that the system s is B working properly. The Self Test T Wavefo orm may be sent by pre essing the key from the MAIN SCREEN N. After the waveform has h been se ent, the ressults will be reflected in the test ressults section of the MA AIN SCREEN N and the PLAYBACK LAST PULS SE SCREEN.
SECULIFE DFBASE This page intentionally left blank.
SECULIFE DFBASE RUNNING A D DEFIBRILLATOR R TEST WAR RNING - CON NNECTIIONS All connec ctions to o patients s must be b remov ved before connecti c ng the Device D Un nder Tes st (D DUT) to the t Analy yzer. A serious s hazard h m may occur if the pattient is connecte c ed when testing w with the Analyzer A r. Do not connect a any leads s from the e patient dire ectly to tthe Analy yzer or DUT.
SECULIFE DFBASE The energy contained in a pulse is determined mathematically based on the fact that the energy is defined as the integral of the power curve. The following formulas describe the basic computation: Energy = E = P dt 2 Power = P = V / R => 2 2 E = V / R dt = V dt / R This computation is implemented digitally by taking timed samples of the voltage every 100 µsec for 100 msec (1000 readings). Each value is then squared and divided by the resistance (50 ohms).
SECULIFE DFBASE DEFIBR RILLATIO ON TEST T The setup for a Defibrillation Te est is depen ndent on the e physical hardware h invvolved. Forr the sake of o this example we will asssume a stan ndard defibrrillator with 5 lead ECG G. WA ARNIN NG This s section n is proviided as a guide to o familiarrize th he user with w the S SECULIF FE DFBASEE. It is no ot inte ended to provide tthe necessary tes st sequen nce for every e De efibrillato or.
SECULIF FE DFBASE NOTE: Revversing the paddles will not cause any damag ge to the un nit or o error in the energy reading. r However, it wiill cause the e polarity off the oscilloscope o e output and the playb back wavefo orm to be inverted. (6 6) Holding the paddles firmly in pla ace, charge e the Defibrrillator and discharge d itt into the SECULIF FE DFBASE. WARN NING O Observe all a precautions n oted by the Defib brillator M Manufactu urer whe en using the Defibrillator..
SECULIFE DFBASE (8) The Status line will change to indicate the various steps as they are being done. (9) At the end of the process the results are continuously displayed in the Test Results section of the MAIN SCREEN. They will remain there until another test is performed, the range is changed or the power is turned off. (10) The user may repeat the playback of the waveform at any time by changing to the PLAYBACK LAST PULSE SCREEN using the key.
SECULIFE DFBASE This page intentionally left blank.
SECULIFE DFBASE CARD DIOVERS SION TE EST A Cardioversion Tesst is simply an energy ttest with sp pecial attention being g iven to the timing. The e SECULIF FE DFBASE continuously c y monitors ffor the R-wa ave timing and a displayys, if possiblle, the delayy between the R-wave e and the pu ulse. In Carrdioversion testing, the e Defibrillatoor is set to deliver d a pecific delay y after the R R-wave.
SECULIFE DF D BASE (4) S Select the de esired ECG G Waveform m and Amplittude to be tested t from the choices on the EC CG W WAVEFORM M SCREEN. NOTE:: The selectted waveforrm must con ntain a QRS S complex. Set the Defib brillator to Synchronize S ed Cardiove ersion mode e. (5) S Place the De efibrillator Paddles P on the SECULIFE DFBASSE contact pplates. The APEX is on the (6) P rig ght and the e STERNUM M is on the lleft.
SECULIFE DFBASE (8) The SECULIFE DFBASE will automatically sense the voltage rise across the internal 50 ohm load and begin taking readings. After the sampling is done (100 ms) the unit will compute and display the results. a. The power pulse is available at the oscilloscope output in real time at 200:1 signal attenuation when in low range and 1000:1 signal attenuation when in high range. b.
SECULIFE DFBASE The user may repeat the playback of the waveform at any time by changing to the PLAYBACK LAST PULSE SCREEN using the key. In this screen the pulse may be viewed in 20 msec segments and paused for review. NOTE: The pulse is sent to the ECG and Paddle outputs at the same time it is being displayed on the screen.
SECULIFE DFBASE CHARG GE TIME E TEST The charrging time of o a Defibrilla ator is noth ing more th han a measu urement of the time req quired for th he Defibrillator to charg ge. It is used d to test the e battery, ch harging circuitry and caapacitor. The SEC CULIFE DFBASE provide es a simple w way to startt and stop the timer. It also record ds the resultts. B W WARNIN NG This section n is provided as a guide to o familiarrize t user with the w the S SECULIF FE DFBASEE.
SECULIFE DF D BASE (5) W While the Pre e-Warning Countdown C iss running, place the Deffibrillator Paaddles on the e SECULIFE DF D BASE conta act plates. Th he APEX is on the right and the ST TERNUM is on o the left. NOTE E: Reversing g the paddle es will not cause c any mage m to the unit or erro or in the energy read ding. Howe ever, it will cause c the po olarity of thee oscope outp put and the playback waveform w to be invertedd.
SECULIFE DFBASE SH OCK AD DVISORY Y AL LGORITH HM TEST T The Shocck Advisory Algorithm A Test works w with the analy ysis and pro ompting funcctions on automatic and semiautomatic Defibrrillators. The ese circuits llook at ECG G waveforms s and promppt the user to o “Shock” orr “No ased on national and international g guidelines.
SECULIFE DFBASE (1) Connect ECG leads to the corresponding universal connector on the SECULIFE DFBASE. The connectors are marked with both the AHA and International color codes. (2) Turn on the SECULIFE DFBASE. (3) The unit will come up in the “High Range Defibrillator” mode. This range is used for normal adult testing. (4) Select the desired AED Waveform and Amplitude to be tested from the choices on the ECG WAVEFORM SCREEN.
SECU ULIFE DFBASE MES SSAGES INPUT OVERLO OAD The “Warning Input Overloa ad Check Range” messsage can diisplay during Defibrillattor testing. The T e should be checked to o see if it should be cha anged to High Range for f the curreent Joule se etting. range LOW BATTERY Y This message m indicates that the batteriies are low and should d be replace ed.
SECULIFE DFBASE SY YSTEM SETUP The SYS STEM SETU UP SCREEN N allows forr the configu uration of th he system ssettings. The e settings can c be Selected using to high hlight the pa arameter an nd using to alloow the editiing of the er. The paramete keys are a used to o edit the setting, then is ussed to accep pt the new setting. T The key can be e used to re eturn to the configuratio on screen w without making a new selection n. The key iss to return to the MAIN N SCREEN.
SECULIFE DFBASE This page intentionally left blank.
SECULIFE DF D BASE POW WER UP SETTING GS The SEC CULIFE DFBASE allows the t user to tailor the se ettings that the unit willl have on Power Up. The B “Power U Up With” parameter in the t System Setup Men nu allows for the selecttion of eithe er Default orr Custom sselections. DEFAUL LT If this opttion is seleccted the following settin ngs will be used u every time the unnit is turned on.
SECULIFE DFBASE This page intentionally left blank.
SECULIFE DF D BASE AUTO A SE EQUENC CE FUNCTION The SEC CULIFE DF FBASE allows s the userr run up to o 50 pre-prrogrammedd sequence es of tests (Auto Sequencces). The tests are configured with h an easy to o use PC program. Eaach test can be configu ured to test Defibrillator, Trranscutaneo ous Pacem maker or bo oth. (For prrogrammingg Auto Seq quences, se ee the Auto Seq quence Prog gramming section).
SECULIFE DFBASE The following table shows the possible test sequence with the details and options that can be selected using the PC program: Test Fields Test Sequence Measures defibrillator Steps Energy Level Limits Defib Energy Tests discharge energy VFIB ECG Output Measures time Do Test? required for defibrillator Energy Level Limits Maximum Energy Test to charge to maximum Max Allowed Charge energy Time Measures Steps Cardioversion Tests Cardioversion Delay Energy Level Limits Steps Tests defibrillato
SECULIFE DFBASE V VIEW MODE The VIEW W MODE allows a the user to look at the test configuration. Each teest setting will w be show wn, as well as the test limiits that iden ntify a valid d or invalid test result. The screenns that are displayed in the VIEW MO ODE are de etermined by b the Auto Sequence selected on n the AUTO O SEQUENC CE SCREEN and its config guration as defined d with h the PC pro ogram.
SECULIFE DFBASE DEFIBRIILLATOR MAXIMUM M ENERGY E T TESTS: Max M Energy Max M Energy Test Limits s DEFIBRIILLATOR CARDIOVE C RSION TES STS: Test T Settingss Test T Limits 64 GMC--I Messtechn nik GmbH
SECULIFE DF D BASE DEFIBRIILLATOR ECG E PERFO ORMANCE E TEST: Waveform W Seelection NOTE: T The individual selected waveformss are not dis splayed in th he VIEW MO ODE.
SECULIFE DFBASE RUN MODE M The RUN N MODE allo ows the userr to run the ttest configurration. The screens s thatt are display yed in the RU UN MODE are determin ned by the Auto A Seque nce selecte ed on the AU UTO SEQU UENCE SCR REEN and its configura ation as deffined with th he PC progrram. Running an Auto Se equence will provide a consistent, guided procedure for ttesting equipment.
SECULIFE DF D BASE The follow wing screen ns may be shown s in th e RUN MODE if all tes st options arre selected: DEFIBRIILLATOR ENERGY E TE ESTS Test Setu up and Actioon Test Passed Test Failed GMC-I Me esstechnik GmbH G 67
SECULIFE DFBASE DEFIBRIILLATOR MAXIMUM ENERGY T ESTS: Te est Setup aand Action Charge Timeer Warning Charge C Timeer Running Resultss 68 GMC--I Messtechn nik GmbH
SECULIFE DFBASE DEFIBRILLATO OR CARDIO OVERSION TESTS: Test Settup and Actiion Test Pass sed DEFIBRILLATO OR ECG PERFORMAN NCE TEST: Waveform m NOTE E: Some tessts, like Perrformance Waveforms, W , do not hav ve quantitative analysees and there efore requirre the user to manuallyy progress through t the test.
SECULIFE DFBASE This page intentionally left blank.
SECULIFE DF D BASE PROG GRAMM MING AUTO SEQ QUENCE ES Auto Sequ uences are programmed p with an easyy to use PC interface. Th his section shhows how to use the Auto o Sequence e configuratio on software. SPL LASH SCREEN The Splassh Screen ide entifies the version v of the e program. This T screen will w be displayyed for 5 sec conds, or unttil the user pressses a key or clicks the sc creen with th e mouse.
SECU ULIFE DFBASE OVE ERVIEW The fo ollowing is a general overrview of the PC Interface e used for Pro ogramming Auto A Sequennces. Each part p of this screen n is describe ed in full deta ail later in this s section.
SECULIFE DFBASE PROGRAM MENU The program menu contains shortcuts to file operations as well as program version information. From the File Menu, you can Save or Load Auto Sequences as well as Clear the History of files that were used. You can also Exit the program from this menu.
SECULIFE DFBASE FILE CONTROL The file control section allows the user to Load and Save Auto Sequence files, greatly expanding the number of pre-programmed sequences from 50 to virtually unlimited. The user also uses the File Control section to Load and Store Auto Sequences on the SECULIFE DFBASE. Dropdown List – This list shows the history of files that have recently been used. This provides quick access for switching between common Auto Sequence files.
SECULIFE DFBASE SEQUENCE SELECTION This section shows a list of all of the names of the Auto Sequences. It is used to select an individual sequence for configuration. Once selected, the configuration window will change to display the settings for the selected sequence.
SECULIFE DFBASE SEQUENCE CONFIGURATION The sequence configuration window displays all of the configuration settings for each Auto Sequence. This sample screen shows a defib test configuration. Sequence Name – This name can be any string of up to 20 standard ASCII characters. NOTE: Not all ASCII characters are valid and will be ignored. Copy Sequence Button – This button opens the Copy Sequence Screen that allows the user to quickly configure similar test sequences.
SECULIFE DFBASE Copy Sequence Screen This function allows the user to quickly configure similar Auto Sequences. Simply select the source and target test sequences, press OK, and the target sequence will be overwritten with the configuration from the source sequence configuration. The Cancel button will exit the window without modifying any configuration settings.
SECULIFE DFBASE Insert Step and Deleting Step Screens The insert and delete functions allow the user to quickly modify an existing test configuration. Deleting a step will move all of the tests following the deleted test up by one step and clear the final step. Inserting a step will shift all following steps down by one step and clear the selected step location. Defibrillator Examples: Pressing OK here would shift steps 4-19 down by one step and insert a blank step at step 4.
SECULIFE DFBASE ECG CONFIGURATION SECTION This section configures the ECG Performance sequence when the ECG Performance Test option is set to YES for a Defibrillator Test. The ECG sequence consists of up to 10 steps. Each step consists of a selected waveform group (Disabled, NSR, AED, Arrhythmia or Performance), output waveform and output amplitude. There is only one ECG sequence for all 50 Auto Sequence tests.
SECULIFE DFBASE This page intentionally left blank.
SECULIFE DF D BASE LIMIITED WA ARRANT TY WARR RANTY: GM MC-I MESST TECHNIK G GMBH WARR RANTS ITS NEW PROD DUCTS TO BE B FREE FR ROM DEFEC CTS IN MAT TERIALS AND WORKM MANSHIP UNDER U THE E SERVICE E FOR WHICH THEY ARE INTEND DED. THIS WARRANTY W IS EFFECT IVE FOR TW WELVE MON NTHS FROM M THE DATE OF SHIPME ENT.
SECU ULIFE DFBASE SPECIF FICATION NS ENER RGY OUTPU UT MEASU UREMENT GEN NERAL M METHOD L LOAD RESISTANCE D DISPLAY RESOLUTIO R ON M MEASUREM MENT TIME E W WINDOW A ABSOLUTE E MAX PEA AK V VOLTAGE P PULSE WID DTH Monop phasic or Biphasic 50 0 ohm +/- 1%, non-iinductive (< <1 µH) 0.1 Joules 100 ms 6000 Volts 100 ms CARDIO OVERSION N D DELAY R RESOLUTIO ON A ACCURACY Y 0 to 6000 ms m 0.
SECULIFE DFBASE ENERGY OUTPUT MEASUREMENT OTHER OSCILLOSCOPE OUTPUT HIGH MEASURE RANGE 1000:1 amplitude-attenuated LOW MEASURE RANGE 200:1 amplitude-attenuated WAVEFORM PLAYBACK OUTPUT LEAD I & PLATES SCREEN 200:1 Time Base Expansion SYNC TIME MEASUREMENTS Starts at peak of TIMING WINDOW each R-wave TEST WAVEFORMS All waveform simulations available DELAY TIME ACCURACY +/- 1 ms CHARGE TIME MEASUREMENT From 0.1 to 99.9 sec ECG NSR 30,40,45,60,80,90,100,120,140,160, 180,200,220,240,260,280,300 BPM +/- 1% 0.
SECULIFE DFBASE ECG AR RRHYTHMIA A SELECTIONS V Ventricular Fibrillation F Atrial Fibrrillation Seccond Degree A-V Block k Righ ht Bundle Branch B Block Prem mature Atria al Contractio on PVC Early E PVC Sta andard PVC R on T Multifoca al PVC Bigem miny Run of 5 PVCs Ve entricular Ta achycardia SHOCK AD DVISORY ALGORITH A M TEST ECG SIG GNALS Asysto ole Coarsse Ventricular Fibrillation Fine e Ventricula ar Fibrillation n Mu ultifocal Ven ntricular Tac chycardia @ 140 BPM Mu ultifocal Ven ntricular Tachyca
SECULIFE DFBASE DATA INPUT/OUTPUTS Parallel Printer Port RS-232C (for computer control) PHYSICAL DISPLAY ENCLOSURE WEIGHT FACE PLATE OPERATING RANGE STORAGE RANGE LCD Graphical 240 X 64 Pixels, Backlit 3.4 x 9.8 x 10.7 Inches (86.4 x 249 x 271.8 mm) Royalite R59 UL Flame Rating 94 V-0 < 5 Lbs (< 2.
SE ECULIFE DFBBASE Prroduct Sup pport If required ple ease contacct: GMC-I M Messtechniik GmbH Productt Support Hotline Phone + +49 911 8602-0 Fax + +49 911 860 02-709 E-Mail ssupport@go ossenmetra awatt.