23.2 Data export to the test device
In order to manage bidirectional data exchange, test devices in the PROFITEST series support the exchange of tree structures, cata-
logs and sequences. These three export categories can be transferred separately or together. This description of the “PROFITEST First
Steps” only deals with simplified data export of the tree structure.
If the electric tree of a facility or machine exists in the IZYTRONIQ, it can also be exported to the connected test device or to a file.
To do this, navigate to the menu “STATIONARY OBJECTS” and select the function “EXPORT”.
This opens an export assistant in which you can define whether export should take place to a file (select folder and specify file name) or
to a test device, and whether the corresponding catalogs and/or sequences should be exported as well.
Once all options are defined, the user selects “EXPORT” to start exporting the object selection.
This opens a window with two parts. The upper half contains the two tree views, while the lower half shows a list of objects selected in
the tree view. Both windows can be used to select the objects for export. Activating the checkboxes in front of the elements in the tree
views is used to make a preselection based on customers and locations. To do this, you can select the objects in the electric tree that
you wish to transfer, which you then restrict through your selection of locations in the location tree.
23.3 Data import from the test device
Similar to data export, users find data import under the function “IMPORT” in the “STATIONARY OBJECTS” menu. This func-
tion is used to synchronize data from a test device or a file with the database in IZYTRONIQ. The import assistant asks whether the object
data should be imported together with the tests. Users can also select whether the catalogs are transferred as well.
To start import, select IMPORT EVERYTHING or SELECT IN IMPORT PREVIEW.
Import takes place automatically if you select IMPORT EVERYTHING and then click on IMPORT.
SELECT IN IMPORT PREVIEW initially shows the data in the IMPORT PREVIEW. The data can then be selected and edited. The
objects are then imported to the IZYTRONIQ database by selecting the function .
A message is shown confirming successful import and any transmission errors.
The conflict manager is activated when there are conflicts between existing and imported data that cannot be resolved automatically.
Take note !
All object data in the test device is deleted if the data is transferred to a
test device.
Take note !
There must be no active measurement on the test device when you start
data export.
Take note !
There must be no active measurement on the test device when you start
data import.