17.2 Structure and operation of the sequence editor
The sequence editor consists of three parts:
1. Top-left window: List of the available sequence steps: This shows the sequence steps that can be used to create the sequence,
depending on the connected devices and type of sequence (IZY/device sequence).
2. Bottom-left window: Configuration of the sequence step: The configuration and parameterization options for the selected step are
offered here.
3. Right-hand window: Display of design progress. This section shows a list with the current sequence progress.
Attention!
The sequences and/or the individual sequence steps may only be cre-
ated by a qualified electrician or under the supervision or instruction of a
qualified electrician by means of the sequence editor.
GMC-I Messtechnik GmbH assumes no liability for sequences created
by unqualified persons!
Attention!
Applying line voltage to the test socket of a SECUTEST or SECULIFE test
instrument as well as a function test are only permitted if the DUT has
passed all safety test steps beforehand! Depending on the safety class of
the DUT, this means that visual inspection, protective conductor resis-
tance and insulation resistance measurement must have been passed
successfully.
Sequence steps
Configuration Design progress