Capacitance Measurement
Dielectric Discharge (DD)
Reference Conditions
Ambient temperature +23 °C ±2K
Relative humidity 40 ... 60%
Measured quantity
frequency 50 Hz ±10 Hz (during voltage
measurement)
Line voltage
waveshape Sinusoidal, deviation between RMS and
rectified value < 1%
Power Supply MERISO PRIME+
Line voltage 207 V
… 253 V / 49 Hz … 61 Hz
or (depending on country-specific design)
108 V
… 132 V / 59 Hz … 61 Hz
Power consumption < 18 VA
Set of batteries NiMH 9.6 V, 3 Ah, charging time 6 hours
Number of measurements
with nominal current
per VDE 0413 700
Power Supply PROFITEST 204HP/HV
Line voltage 207 V
… 253 V / 49 Hz … 61 Hz
Power consumption PROFITEST 204HP-2.5kV: max. 700 VA
PROFITEST 204HV-5.4kV: max. 100 VA
Electrical Safety
Standards IEC 61010-1:2001
EN 61010-1:2002
VDE regulations VDE 0411 Part 1:2002
Pollution degree 2
Protection IP 40
– MERISO PRIME+
Measuring category
Insulation measurement – 5000 V DC –
no overvoltage
Voltage measurement – 1000 V – CAT II
Safety class II
– PROFITEST 204HP/HV
Safety class I
Safety shutdown if instrument overheats
Fuses mains: F 3.15 / 250
Display
Range
Measuring
Range
Test Voltage Intrinsic Uncertainty Measuring
Uncertainty
0.00
…
10.0
μ
F 0.10
…
5.00
μ
F
100…450 V ±(10 % rdg. + 5 d) ± (15% rdg. + 8 d)
500…5 kV ± (5 % rdg. + 5 d) ± (10% rdg. + 8 d)
Limit
DD 0.10 ... 2.00 ... 9.80