Operation Manual
GMC-I Messtechnik GmbH 47
4.6.1 Setting Test Parameters
The discharge time during which the voltage value must drop to a
non-hazardous value of less than 60 V can be specified as a limit
value between 1.0 and 9.00 seconds. The following message
appears in order to indicate that the limit has been reached:
“residual voltage less than 60 V!”.
Set Discharge Time
Test duration is always set to continuous measurement in this case
because residual voltage testing is triggered automatically and
voltage measurement is always active for safety reasons.
For resetting parameter values to default setting, see page 22.
4.6.2 Starting the Test
START
STORE
Press briefly: Save results
Press & hold: Save results and activate the entry field:
A description of the measurement with up to 15
characters can be entered to the information field
at the bottom. See chapter 5.1.1 regarding entry.