Operation Manual

44 GMC-I Messtechnik GmbH
4.5 Capacitance Measurement and
Determination of Dielectric Discharge (optional)
Capacitance measurement is performed with adjustable test volt-
age by charging with constant current. The measurement is per-
formed with automatic range selection. One of the nominal test
voltages (U
nom
), or a freely selectable test voltage (U
var
) can be
used.
Dielectric Discharge Test (DD)
This test was developed for use with large machines and is per-
formed, as its name indicates, during discharge. It designates the
ratio of the discharge current after one minute and the product of
voltage by capacitance:
DD values > 4: Troubleshooting is required
DD values = 2 4 Maintenance is recommended
DD values < 2 DUT is OK, no immediate
action is required.
Application: Determination of the magnitude of absorbed contami-
nation or moisture
TEST
Select the Test
MENU
Open the Submenu
4.5.1 Setting Test Parameters
Set Parameters
For resetting parameter values to default setting, see page 22.
DD
I
Disch
1min()
UC
------------------------------
1000=