4.4.1 Setting Test Parameters
Set Parameters
Significance of Parameters with Variable and Fixed Test Voltage
Test voltage U
nom
: Test voltage in fixed steps
(100/250/500 V, 1.00/1.50/2.00/2.50/5.00 kV)
U
var
: Variable test voltage (100 V … 5.00 kV)
U
ins
Test voltage (see fixed steps above)
Rise Time after which maximum test voltage will be reached
(5 s … 300 s)
Test
duration Test duration at maximum test voltage
(AUTO, 1 s … 120 s,
>>>> (continuous, burn mode))
For resetting parameter values to default setting, see page 22.
4.4.2 Starting the Test
START
STORE
Press briefly: Save results
Press & hold: Save results and activate the entry field:
A description of the measurement with up to 15
characters can be entered to the information field
at the bottom. See chapter 5.1.1 regarding entry.