Operation Manual

38 GMC-I Messtechnik GmbH
Saving Measured Values
After the test sequence has been completed, the last measured
U
T
and I
T
values remain at the display. Results can be saved to
memory with or without a comment. An automatically generated
consecutive number is assigned to the results for allocation to the
respective test.
A description with up to 15 characters can be entered for the
measurement to the field at the bottom left. See chapter 5.1.1 on
page 49 regarding entry.
The measured values at the display are overwritten when the next
test is executed.
STORE
Press briefly: Save results
Press & hold: Save results and initialize entry field
In the continuous operation mode (test duration set to >>>>>>>),
no values are saved to memory for U
T
and I
T
.
Aborting the High-Voltage Test Before Completion
The high-voltage test can be aborted prematurely at any time by
releasing the trigger at the high-voltage switching pistol.
If testing is aborted by releasing the trigger at the high-voltage
switching pistol before the selected test duration has elapsed,
displayed U
T
and I
T
values may deviate from the last measured
values as a result.
Note
Breakdown Voltage
If interrupting current I
MAX
is exceeded
before the selected test voltage has been
reached, test voltage U
T
and current I
MAX
measured at
this point in time appear at the display preceded by the
“>” symbol, and are saved to memory. The appropriate
symbol is displayed simultaneously.