4.2.2 Setting Test Parameters for High-Voltage Testing
The following parameters can be set in the “standby” mode with
illuminated green signal lamps:
Test v. AC-HV
U
AC.MAX.
Max. test voltage, final value after T
R
has elapsed
Setting limits: HP-2.5kV: 250 V … 2.50 kV
HV-5.4kV: 650 V … 5.35 kV
I
MAX.
: Max. allowable current before
high-voltage is switched off
Setting limits: HP-2.5kV: 10.0 … 250 mA
HV-5.4kV: 0.50 … 10.0 mA
The pulse mode symbol >>>>>>> can be selected with the
key at less than 10.0 or 0.50 mA.
T
D
: Test duration at maximum test voltage U
AC.MAX.
(not including rise time T
R
)
Setting limits: 1.00 … 120 s, the continuous
operation symbol >>>>>>> can
be selected at under 1.00 s
with the key.
T
R
: Rise time: time, during which test voltage
rises to the selected value U
AC.MAX.
Setting limits: 100 ms … 99.9 s
MENU
Select Parameter
Set Value
Test
voltage
Rise
Test duration Decline*
Tes t pe r io d
Time
* constant = 0.1 s