Operation Manual

GMC-I Messtechnik GmbH 33
4.1.5 Measurements with the Guard Cable (feature G1)
The measurement of very high resistances necessitates extremely
minimal measuring current and may be rendered problematic as a
result of influences such as electromagnetic fields, humidity or
surface current. An accurate test set-up is thus absolutely essen-
tial.
A guard cable must be used for measurements within a range of
100 GΩ (10 GΩ) 1TΩ, in order to prevent surface current from
distorting measurement results. The guard rings prevent current
at the surface of the insulation material from flowing from the
+measurement cable to the measurement cable, instead of
through the insulation material itself.
Ð Insert the plug from the guard cable into the appropriate jack in
the test instrument.
Ð Plug the alligator clip onto the guard cable test probe.
Ð Connect the alligator clip to the guard ring between the two
measuring points at the insulation material under test.
Ð Refer to chapter 4.1.2 on page 31 regarding the measuring se-
quence
Note
The following materials can be used as guard rings: alu-
minum foil, copper foil or metallic hose clamps.
+Measurement Cable
Measurement Cable
Guard Cable
Conductor
Insulation
Material
Guard Rings
Contact Ring