Operation Manual

32 GMC-I Messtechnik GmbH
Statistical analysis
Press the CURSOR DOWN or CURSOR UP keys to return to the previous
LCD display.
Note
If the measuring range is exceeded (overflow of display
range), no statistical analysis will be performed. In this
case, the individual measured values are listed, but no
value for ΔR
MAX
is calculated.
STORE
Press briefly: Save results
Press & hold: Save results and activate the entry field:
A description of the measurement with up to 15
characters can be entered to the information field
at the bottom. See chapter 5.1.1 regarding entry.
4.1.4 Data logger function (min log)(as of firmware version AH)
MENU
Select Parameter
If you choose time parameter setting "min log" from the menu,
exactly 30 measured values are automatically stored during the
selected test duration at time intervals between 4 seconds and
2 minutes. This makes data logger operation possible for measur-
ing durations between 2 and 60 minutes. In this case, the residual
time of the current measurement is not indicated in seconds, but
in minutes. While the test is being performed, icon “Stored” is
briefly shown in the display and the yellow signal lamp “Testing ...
disappears briefly each time a measured value is being stored. If
the signal generator is activated, an acoustic signal sounds addi-
tionally.
In this function, the current test time is stored for each measure-
ment instead of the parameter for the test duration to ensure the
temporal correlation between measured value and time of mea-
surement.
Note
Before the start of each test with data logger as
described in chapter 5.1, select a new device under test
to avoid the measured values being stored successively
within the same DUT.
This considerably facilitates the subsequent evaluation of
the stored measured values.