Datasheet
GMC-I Messtechnik GmbH 3
METRISO INTRO, BASE, TECH
High-Precision
Insulation, Low Resistance and Voltage Measurement Instrument
Reference Conditions
Reference
temperature + 23 C 3 K
Relative humidity 40 ... 75%
Measured quantity
frequency 45 Hz ... 65 Hz
Measured quantity
waveshape Sine, deviation between TRMS and recti-
fied value < 1%
Battery voltage 9.5 V 0.1 V
Test resistor 10 M1%
Electrical Safety
Protection class II
Pollution degree 2
Measuring category METRISO INTRO/TECH:
CAT II 1000 V / CAT III 600 V / CAT IV 300 V
METRISO BASE
:
CAT III 600 V / CAT IV 300 V
Fuses
Fuse link FF315mA/1000V, effective in all resis-
tance measuring ranges, 1 additional
replacement fuse in the battery compartment
Elektronic fuse for protecting low-resistance and resis-
tance measurement R
LO
and R
(not METRISO G500MM (M550K))
Ambient Conditions
Accuracy
temperature range 0 ... +40 C
Operating
temperature –10 ... +50 C
Storage temp. range –25 ... +70 C (without batteries)
Relative humidity Up to 75% (max. 85% during storage/
transport), no condensation allowed
Elevation Max. 2000 m
Calibration interval 1 year (recommended)
Electromagnetic Compatibility (EMC)
Interference emission EN 61326-1:2006 class B
Interference immunity EN 61326-1:2006
Displays
Digital Display With additional bar graphs or pointer
depending on selection with a.disp
parameter, backlit (transflective);
leading zeros can be suppressed at the
digital display depending on selection with
0.disp parameter;
overranging indicated with 0L at display;
dimensions: 65 x 36 mm
Cable resistance
If measurement results for the two direc-
tions of current flow (polarity reversal) differ
by more than 10% (this corresponds to
typical measuring error for the instru-
ments), both measured values are dis-
played next to each other with reduced
resolution.
Limit LED LED lights up red to indicate an exceeded
limit value
LED lights up green to indicate adherence
to the limit value
LED LED lights up red to indicate:
– the presence of an external voltage
bevore insulation testing
(U > 50 V)
with the device switched on or off
(device switched off not with M550N)
– the presence of the test/measuring voltage
during (insulation) measurement (U > 50 V)
the presence of a residual voltage
after insulation testing (U > 50 V)
with the device switched on or off
at LCD Detection of external voltage at the LCD
with the device switched on where U DC >
50 V and U AC > 40 V (50 Hz) for all mea-
suring functions
Mechanical Design
Dimensions 225 x 130 x 140 mm
Weight Approx. 1.4 kg with batteries
Protection Housing: IP 52, measurement cables and
connectors: IP 40 per DIN VDE 0470, part
1 / EN 60529, housing category 2
Extract from table on the meaning of IP codes
IP XY
(1
st
digit X)
Protection Against Foreign
Object Entry
IP XY
(2
nd
digit Y)
Protection Against
Penetration by Water
2 12.5 mm dia. 2 Dripping (at 15° angle)
3 2.5 mm dia. 3 Spraying water
4 1.0 mm dia. 4 Splashing water
5 Dust protected 5 Jet-water
6 Dust-proof 6 Powerful water jets
M Vk
TRMS
AC
DC
Rlo>10%
MAN
DATA
ZERO
G
ISO
2500V
1000V
Tk
1
0
0
k
1
G
1
0
k
1
M
1
0
G
0
1
0
0
0
V
1
2
0
0
G
1
0
0
M
1
0
M
!